Improvement in Metal-on-Metal Bearing of Hip Implants with Advanced Stellite Alloys

Abstract

dc:description.abstract

CoCrMo alloy Stellite 21 has been used as hip implant material decades. However, its limited metal-on-metal bearing has resulted in loosening in hip implants. Three advanced alloys, Stellite 720, Stellite 21 with increased 10 wt%Cr (named modified Stellite 21) and Stellite 21 with additional 0.4 ~ 0.6 wt%CrN (named nitrided Stellite 21), are proposed in this research to improve the bearing. The wear test is performed on a pin-on-disc tribometer. The corrosion tests are conducted in the simulated human body environment. The results demonstrate that the proposed alloys all exhibit better wear resistance than the conventional Stellite 21. The corrosion resistance of modified Stellite 21 is much better than that of Stellite 21 while Stellite 720 and nitrided Stellite 21 are worse than Stellite 21 in corrosion resistance. Modified Stellite 21 with combined superior wear and corrosion resistance properties is highly recommended.

Degree

thesis:*
Name thesis:degree_name
Master of Applied Science (M.App.Sc.)
Level thesis:degree_level
Master's
Discipline thesis:degree_discipline
Engineering, Materials
Grantor dc:publisher
Carleton University
Year dc:date.issued
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Hu, Peishan

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • Copyright © 2013 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner.
Language dc:language.iso
en

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Carleton University
Base URL
carleton.scholaris.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Hu, Peishan. Improvement in Metal-on-Metal Bearing of Hip Implants with Advanced Stellite Alloys. Master's thesis, Carleton University, 2013. https://hdl.handle.net/20.500.14718/32830