Back to search

Carleton University

Kilo-Voltage X-Ray Correction Factors for In-Water Measurements Based on TG-61

Abstract

dc:description.abstract

For x-ray tube potentials 100 kV, the AAPM TG-61 protocol for 40-300 kV x-ray beam dosimetry in radiotherapy recommends an in-water measurement which is based on ionization chambers calibrated in air in terms of air kerma. The variation of the overall correction factor (PQ;cham) and its components (known as corrections for the change in the chamber response due to the change in the spectrum distribution in phantom compared to that used for the calibration free in-air (PE;), displacement of water by the ionization chamber (Pdis) and correction due to the stem for in-water measurement relative to the free in-air calibration (Pstem;w)) were studied with depth and field size for 6 different beam qualities in the orthovoltage x-ray range (100 kV tube potential 300 kV). The NE2571 ion chamber, Exradin A16 ion chamber, and Exradin W1 plastic scintillator were considered in this thesis. The last two detectors are not included in the TG-61 protocol.

Degree

thesis:*
Name thesis:degree_name
Master of Science (M.Sc.)
Level thesis:degree_level
Master's
Discipline thesis:degree_discipline
Physics
Grantor dc:publisher
Carleton University
Year dc:date.issued
2016

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Sherafati, Nima

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • Copyright © 2016 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner.
Language dc:language.iso
en

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Carleton University
Base URL
carleton.scholaris.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Sherafati, Nima. Kilo-Voltage X-Ray Correction Factors for In-Water Measurements Based on TG-61. Master's thesis, Carleton University, 2016. https://hdl.handle.net/20.500.14718/39505