Analysis of Compressively Sensed Electrocardiogram for Detection of Atrial Fibrillation

Abstract

dc:description.abstract

Cardiovascular diseases (CVDs) are one of the leading causes of death in the world. Long-term ambulatory electrocardiogram (ECG) monitoring enables timely medical interventions. Effective and efficient ambulatory ECG monitoring can be supported by compressive sensing, which can reduce battery consumption; however, the reconstruction of compressively sensed ECG could be a computationally intensive operation. Therefore, detection of CVDs in compressively sensed ECG is highly desirable. This thesis proposes a three-stage system based on machine learning for detecting atrial fibrillation (AF) in compressively sensed ECG while reducing false alarms (i.e., false positives) due to contamination and reducing reconstruction of the ECG. Stage 1 uses a novel signal quality index (SQI) to represent the quality of the compressively sensed ECG on a continuous scale and reject low-quality ECG. Stage 2 classifies the ECG as either AF or normal and generates an associated confidence score. Stage 3 selectively reconstructs the ECG when the Stage 2 confidence score is unacceptable and classifies the reconstructed ECG as either AF or normal. Clean ECG from Long-Term AF Database was artificially corrupted with simulated motion artifact to pre-set signal-to-noise ratios. The corrupted ECG was compressively sensed at 50% and 75% compression ratios (CRs). The system was evaluated using average precision (AP), the area under the curve (AUC) of the receiver operator characteristic curve, false-positive rate (FPR), and true positive rate (TPR). The system was optimized to maximize the AP and minimize ECG rejection and reconstruction ratios. The optimized system for 50% CR had 0.72 AP, 0.63 AUC, 0.38 rejection ratio, and 0.38 reconstruction ratio. The optimized system for 75% CR had 0.72 AP, 0.63 AUC, 0.40 rejection ratio, and 0.35 reconstruction ratio. The optimized system reduced FPR by 0.37 for both CRs compared to Stage 2 alone. At a fixed FPR of 0.10, the TPR of the proposed system was 0.43 and 0.42 for 50% and 75% CRs, respectively, an improvement of 0.10 and 0.14 over Stage 2. The proposed system reduced the probability of false alarms (FPR) and detected AF in compressively sensed ECG (improved TPR) while being computationally efficient (low reconstruction ratio).

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy (Ph.D.)
Level thesis:degree_level
Doctoral
Discipline thesis:degree_discipline
Engineering, Electrical and Computer
Grantor dc:publisher
Carleton University
Year dc:date.issued
2022

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Abdelazez, Mohamed Hazem

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • Copyright © 2022 the author(s). Theses may be used for non-commercial research, educational, or related academic purposes only. Such uses include personal study, research, scholarship, and teaching. Theses may only be shared by linking to Carleton University Institutional Repository and no part may be used without proper attribution to the author. No part may be used for commercial purposes directly or indirectly via a for-profit platform; no adaptation or derivative works are permitted without consent from the copyright owner.
Language dc:language.iso
en

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
Carleton University
Base URL
carleton.scholaris.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Abdelazez, Mohamed Hazem. Analysis of Compressively Sensed Electrocardiogram for Detection of Atrial Fibrillation. Doctoral thesis, Carleton University, 2022. https://hdl.handle.net/20.500.14718/42367