Abstract
dc:description.abstractThe response of an Electromagnetic Flowmeter to a secondary, nonconducting phase in a continuous conducting phase is investigated. Experiments are carried out on a commercial 2" Foxboro flowmeter with sinusoidal mains field excitation and point electrodes. The main aspects which are looked for are the change in flow signal and transformer signal. A mathematical model is developed to simulate the effect of the nonconducting phase in. the form of single spherical particles. The model also allows one to alter the electrode geometry. A second flowmeter is designed to incorporate an improved electrode and coil geometry as well as a newly designed signal processing unit. In a series of experiments, the predictions of the theoretical model have been mostly verified.
Degree
thesis:*- Grantor dc:publisher
- Cranfield University
- Year dc:date.issued
- 1993
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Krafft, Rainer
- Advisors dc:contributor.advisor
-
- Sanderson, M. L.
- Hemp, J.
Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Repository record dc:identifier.uri
- http://dspace.lib.cranfield.ac.uk/handle/1826/4637