Back to search

Dalhousie University

Determining Mean Corpuscular Hemoglobin from a Drop of Blood using Contact Optical Microscopy

Abstract

dc:description.abstract

Contact optical microscopy (COM) is a novel and highly compact imaging technology that has the potential to deliver a point-of-care complete blood count, which could strongly impact many fields of medicine. It was hypothesized that by using COM, Mean Corpuscular Hemoglobin (MCH) can be measured as a component of a CBC, with clinically acceptable accuracy and precision, from a drop of blood. A physical COM system, a microspectrophotometry equation, and a computer-vision algorithm were developed to extract the MCH from COM images of films of blood containing intact red blood cells. The MCH of 15 human blood samples from different subjects was measured using both a standard-of-practice Sysmex XN-10 machine and the COM system. A coefficient of variation of 4% was obtained, surpassing the clinical target of 5%. The results showed that MCH can be measured precisely using COM-based microspectrophotometry and suggest that there is potential for further precision gains.

Degree

thesis:*
Department dc:contributor.department
Department of Biomedical Engineering
Year dc:date.issued
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Macaulay, Hershel

Subjects

dc:subject × 10

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/10222/73948

Chain of custody

source
Harvested from
Dalhousie University
Base URL
dal.scholaris.ca/server/oai/request
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Macaulay, Hershel. Determining Mean Corpuscular Hemoglobin from a Drop of Blood using Contact Optical Microscopy. 2018. https://hdl.handle.net/10222/73948