Back to search

Dublin City University

Remote sensing of micro pollutants in water using a fibre optic analysis system

Abstract

dc:description.abstract

The real-time detection of phenols in groundwater using a fibre optic fluorescent sensor system is discussed. Two sensor designs are discussed, modelled and experimentally evaluated. One, consisting of a single source fibre and nine ‘return’ fibres used with a deuterium lamp source, double monochromator and photomultiplier detector is compared to a dual fibre device which is excited by a frequency quadrupled Nd-Yag laser source, narrow pass band optical filter and photomultiplier detector. The latter system has a reference channel to compensate for temporal variations in laser output optical power. Phenol resolution of less than 1 ppb was achieved with the laser based sensor operating over a source-sample distance of 15 metres.

Degree

thesis:*
Name dc:type.qualificationname
msc
Level dc:type.qualificationlevel
masters
Grantor dc:publisher.institution
Dublin City University
Year dc:date.issued
1993

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Higgins, Kieran

Rights

Language dc:language
en

Chain of custody

source
Harvested from
Dublin City University
Base URL
doras.dcu.ie/cgi/oai2
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Higgins, Kieran. Remote sensing of micro pollutants in water using a fibre optic analysis system. masters thesis, Dublin City University, 1993.