title="Search in OATD">Florida Atlantic University
CLINICAL VALIDATION OF DELIVERY ANALYSIS SOFTWARE FOR PATIENT-SPECIFIC QUALITY ASSURANCE OF RADIXACT (TOMOTHERAPY) TREATMENT
Abstract
dc:description.abstractThe primary objective of this study was to evaluate the patient-specific quality assurance (PSQA) performance for various disease sites treated with the Radixact machine using Gamma (γ) passing rate analysis (%GP) as per AAPM TG-218. This study analyzed 22 PSQA plans executed on a Radixact machine using ArcCheck phantom and delivery analysis (DA) software. The %GP was calculated using a 3% dose difference (DD), a 2 mm distance-to-agreement (DTA), and a 10% threshold. A passing criterion of ≥95% was applied for PSQA evaluation. The %GP was analyzed for all sites. The %GP (2D) on the ArcCheck phantom mean γ passing rate with variability was (97.12±2.63%), while the %GP (3D) rate for DA was 100%. For DA, the %GP was 100% because it only considers the MLC-LOTs patterns, not plan parameters, to compute the delivered versus planned dose. Investigations should be required to recalibrate the output of the MVCT diode array.
Degree
thesis:*- Name thesis:degree_name
- Professional Science Masters (PSM)
- Level thesis:degree_level
- Thesis
- Discipline thesis:degree_discipline
- Physics
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Pama, Gagandeep
Subjects
dc:subject × 11Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.fau.edu/etd_general/44