Back to search

title="Search in OATD">Florida Atlantic University

Machined brass skin collimation with variable thickness for electron therapy

Abstract

dc:description.abstract

Skin collimation in electron therapy ensures sharper penumbra and maximal protection to adjacent critical structures. It also provides a better clinical dose to the target and avoids recurrences at the periphery. The thickness of the electron skin collimation must be adequate for shielding purposes, not too thick to cause discomfort to the patient and be conformal to the skin. This study assessed the clinical potential of machined brass skin collimation with variable thickness. Brass transmission factors for 6, 9, and 12 MeV electron beams were measured and used to determine the skin collimation clinically acceptable thickness. Dosimetric performance of the variable thickness skin collimation was evaluated for 9 MeV electrons within a rectilinear water-equivalent phantom and a water-filled head phantom. Results showed the variable thickness skin collimation is dosimetrically equivalent to the uniform thickness collimation. Favorable dosimetric advantages for brass skin collimation for small electron fields were achieved.

Degree

thesis:*
Level thesis:degree_level
Thesis

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Gomez, Facenda Alianna

Subjects

dc:subject × 3

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.fau.edu/etd_general/771

Chain of custody

source
Harvested from
title="Search in OATD">Florida Atlantic University
Base URL
digitalcommons.fau.edu/do/oai/
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Gomez, Facenda Alianna. Machined brass skin collimation with variable thickness for electron therapy. Thesis thesis, https://digitalcommons.fau.edu/etd_general/771