title="Search in OATD">Florida Atlantic University
Design and construction of a high-sensitivity capacitance-based quartz dilatometer
Abstract
dc:description.abstractA high sensitivity capacitance dilatometer constructed solely of fused quartz was designed to study phase transitions of sub-millimeter sized samples from 4 to 300 K. The minimal thermal expansion contribution from quartz reduced the overall magnitude of the background by a factor of 10. As a consequence, the increased resolution of the data provided greater sensitivity than has been capable with metallic capacitance dilatometers. A one millimeter long single crystal of La1.67Sr0.33NiO 4 was analyzed from 80 to 300 K. Our highly reproducible measurements reveal phase transitions at similar temperatures to those observed with magnetization and specific heat measurement techniques.
Degree
thesis:*- Level thesis:degree_level
- Thesis
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Timmins, George Emerson
Subjects
dc:subject × 4Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.fau.edu/etd_general_5/367