Back to search

title="Search in OATD">Florida Atlantic University

Measurements of output factors for small photon fields up to 10 cm x 10 cm

Abstract

dc:description.abstract

Field output factors (OF) for photon beams from a 6 MV medical accelerator were measured using five different detectors in a scanning water phantom. The measurements were taken for square field sizes of integral widths ranging from 1 cm to 10 cm for two reference source-to-surface distances (SSD) and depths in water. For the diode detectors, square field widths as small as 2.5 mm were also studied. The photon beams were collimated by using either the jaws or the multileaf collimators. Measured OFs are found to depend upon the field size, SSD, depth and also upon the type of beam collimation, size and type of detector used. For field sizes larger than 3 cm x 3 cm, the OF measurements agree to within 1% or less. The largest variation in OF occurs for jawsshaped field of size 1 cm x 1cm, where a difference of more than 18% is observed.

Degree

thesis:*
Level thesis:degree_level
Thesis

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Bacala, Angelina

Subjects

dc:subject × 4

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.fau.edu/etd_general_6/817

Chain of custody

source
Harvested from
title="Search in OATD">Florida Atlantic University
Base URL
digitalcommons.fau.edu/do/oai/
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Bacala, Angelina. Measurements of output factors for small photon fields up to 10 cm x 10 cm. Thesis thesis, https://digitalcommons.fau.edu/etd_general_6/817