title="Search in OATD">Florida Atlantic University
The importance of immobilization and localization of gynecological applicators in high dose rate brachytherapy treatments
Abstract
dc:description.abstractIntracavitary high dose rate (HDR) brachytherapy is a form of radiation therapy generally in which a post-surgical tissue margin is treated. The dose gradient of HDR brachytherapy is very steep, and thus small displacements of the applicator, even as small as 1 mm, could potentially cause significant variations of dose which could result in undesired side effects such as overdose of a critical organ. In this retrospective dosimetric study, the variation of dose due to various small range motions of gynecological applicators is investigated. The results show that the implementation of additional immobilization and localization devices along with other safety measures needs to be further investigated.
Degree
thesis:*- Level thesis:degree_level
- Thesis
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Hyvärinen, Mikko
Subjects
dc:subject × 7Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.fau.edu/etd_general_6/1340