Freie Universität Berlin
Reduction of claustrophobia with short-bore versus open magnetic resonance imaging
Abstract
dc:description.abstractBackground: Claustrophobia is a common problem precluding magnetic resonance imaging (MRI). It occurs in up to 35% of all MR examinations. Between 1% and 15% of all examinations cannot be performed or conscious sedation is required to complete the scan due to claustrophobia. A more open MR scanner configuration is one approach to reduce claustrophobic reactions. The purpose of the present study was to assess whether a short-bore or an open MR scanner is superior in alleviating claustrophobia. Methods: Institutional review board approval and patient informed consent were obtained to compare short-bore versus open MR. From June 2008 to August 2009, 174 patients (139 women; mean age = 53.1 [SD 12.8]) with an overall mean score of 2.4 (SD 0.7, range 0 to 4) on the Claustrophobia Questionnaire (CLQ) and a clinical indication for imaging, were randomly assigned to receive evaluation by open or by short-bore MR. The primary outcomes were incomplete MR examinations due to a claustrophobic event. Follow-up was conducted 7 months after MR imaging. The primary analysis was performed according to the intention-to-treat strategy. Results: With 33 claustrophobic events in the short-bore group (39% [95% confidence interval (CI) 28% to 50%) versus 23 in the open scanner group (26% [95% CI 18% to 37%]; P = 0.08) the difference was not significant. The imaging duration of patients who completed their MR examination was significantly shorter in the short-bore group (31.7 min [SD 13.2] vs. 42.1 min [SD 19.5]; P = 0.001). Patients with an event were in the examination room for 3.8 min (SD 4.4) in the short-bore and for 8.5 min (SD 7) in the open group (P = 0.004). This was due to an earlier occurrence of events in the short-bore group. The CLQ suffocation subscale was significantly associated with the occurrence of claustrophobic events (P = 0.003). New findings that explained symptoms were found in 69% of MR examinations and led to changes in medical treatment in 47% and surgery in 10% of patients. After 7 months, perceived claustrophobia increased in 32% of patients with events versus in only 11% of patients without events (P = 0.004). Conclusions: Claustrophobic event rates were disappointingly high at over 25% despite the modern design of both short-bore and open MR in this high-risk group. Further developments to create a more patient-centered MR scanner environment are needed to make this important diagnostic test available to all patients.
Degree
thesis:*- Department dc:contributor.department
- Charité - Universitätsmedizin Berlin
- Year dc:date.issued
- 2014
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Enders, Judith
- Committee member dc:contributor.committeemember
-
- N.N.
Subjects
dc:subject × 9Rights
dc:rights.accessrights- Statement dc:rights.accessrights
-
- open access
- Licence dc:rights.uri
- Language dc:language
- eng, ger
Identifiers
dc:identifier.*- URN
- urn:nbn:de:kobv:188-fudissthesis000000096619-7
- Identifier URI
- http://dx.doi.org/10.17169/refubium-10041
- Dc Identifier Other
-
FUDISS_thesis_000000096619
FUDISS_derivate_000000015132