Georgia Institute of Technology
Dosimetric characterization of elongated brachytherapy sources using Monte Carlo methods
Abstract
dc:description.abstractCurrent brachytherapy treatment planning systems are unable to accurately calculate dose distributions in the vicinity of brachytherapy sources having active lengths much greater than 5 mm. While low dose-rate ¹³⁷Cs sources are dosimetrically characterized using antiquated along-away tables with simple linear-linear interpolation errors in dose calculation exceeding 30% occur due to algorithm inadequacy. The method presented in this thesis permits dosimetric characterization of elongated brachytherapy sources with active lengths 0 < L < 10 cm for implementation on an FDA-approved clinical TPS. Low- and high-energy photon-emitting sources of Pd-103 and Ir-192, respectively, were examined.
Degree
thesis:*- Department dc:contributor.department
- Mechanical Engineering
- Grantor dc:publisher
- Georgia Institute of Technology
- Year dc:date.issued
- 2010
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Bannon, Elizabeth
- Advisors dc:contributor.advisor
-
- Rivard, Mark J.
- Wang, C.-K. Chris
- Committee member dc:contributor.committeemember
-
- Cho, Sang Hyun
Subjects
dc:subject × 4Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/1853/33923