Automated Extraction of Subdural Grid Electrodes from Post-Implant MRI Scans for Epilepsy Surgery

Abstract

dc:description.abstract

The objective of the current research was to develop an automated algorithm with no or little user assistance for extraction of Subdural Grid Electrodes (SGE) from post-implant MRI scans for epilepsy surgery. The algorithm utilizes the knowledge about the artifacts created by Subdural Electrodes (SE) in MRI scans. Also the algorithm does not only extract individual electrodes, but it also extracts them as a SGE structures. Information about the number and type of implanted electrodes is recorded during the surgery [1]. This information is used to reduce the search space and produce better results. Currently, the extraction of SGE from post-implant MRI scans is performed manually by a technologist [1, 2, 3]. It is a time-consuming process, requiring on average a few hours, depending on the number of implanted SE. In addition, the process does not conserve the geometry of the structures, since electrodes are identified individually. Usually SGE extraction is complicated by nearby artifacts, making manual extraction a non-trivial task that requires a good visualization of 3D space and orientation of SGE in it. Currently, most of the technologists use 2D slice viewers for extraction of SGE from 3D MRI scans. There is no commercial software to perform the automated extraction task. The only algorithm suggested in the literature is [4]. The goal of the proposed algorithm is to improve the performance of the algorithm in [4]. As a goal, the proposed algorithm performs extraction of SGE not only for individual electrodes, but by applying geometric constraints on SGE.

Degree

thesis:*
Department dc:contributor.department
Electrical and Computer Engineering
Grantor dc:publisher
Georgia Institute of Technology
Year dc:date.issued
2004

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Pozdin, Maksym O.
Advisor dc:contributor.advisor
  • Skrinjar, Oskar
Committee members dc:contributor.committeemember
  • Anthony Yezzi
  • John Oshinski

Subjects

dc:subject × 9

Rights

Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1853/4979

Chain of custody

source
Harvested from
Georgia Tech
Base URL
repository.gatech.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Pozdin, Maksym O.. Automated Extraction of Subdural Grid Electrodes from Post-Implant MRI Scans for Epilepsy Surgery. Georgia Institute of Technology, 2004. http://hdl.handle.net/1853/4979