Abstract

dc:description.abstract

A number of clinical, in vitro and computational studies have shown the potential for thromboembolic complications in bileaflet mechanical heart valves (BMHV), primarily due to the complex and unsteady flows in the valve hinges. These studies have focused on quantitative and qualitative parameters such as velocity magnitude, turbulent shear stresses, vortex formation and platelet activation to identify potential for blood damage. However, experimental characterization of the whole flow fields within the valve hinges has not yet been conducted. This information can be utilized to investigate instantaneous damage to blood elements and also to validate numerical studies focusing on the hinge’s complex fluid dynamics. The objective of this study was therefore to develop a high-resolution imaging system to characterize the flow fields and global velocity maps in a BMHV hinge. Subsequently, the present study investigated the effect of hinge gap width on flow fields in a St. Jude Medical BMHV. The results from this study suggest that the BMHV hinge design is a delicate balance between reduction of fluid shear stresses and areas of flow stasis during leakage flow, and needs to be optimized to ensure minimal thromboembolic complications. Overall, the current study demonstrates the ability of high-resolution Micro Particle Image Velocimetry to assess the fluid flow fields within the hinges of bileaflet mechanical heart valves, which can be extended to investigate micro-scale flow domains in critical regions of other cardiovascular devices to assess their blood damage potential.

Degree

thesis:*
Level thesis:degree_level
Masters
Department dc:contributor.department
Mechanical Engineering
Grantor dc:publisher
Georgia Institute of Technology
Year dc:date.issued
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Jun, Brian H.
Advisor dc:contributor.advisor
  • Yoganathan, Ajit P.
Committee members dc:contributor.committeemember
  • Dixon, Brandon
  • Giddens, Don P.

Subjects

dc:subject × 6

Rights

Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1853/53380

Chain of custody

source
Harvested from
Georgia Tech
Base URL
repository.gatech.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Jun, Brian H.. In vitro micro particle image velocimetry measurements in the hinge region of a bileaflet mechanical heart valve. Masters thesis, Georgia Institute of Technology, 2014. http://hdl.handle.net/1853/53380