Back to search

Georgia Institute of Technology

Experimental validation of a collision avoidance software in radiation therapy

Abstract

dc:description.abstract

Cancer is one of the deadliest diseases in the United States. Advancements in the field of radiation therapy, like the development of image-guided radiation therapy, intensity modulation, and volumetric-modulated arc therapy, has increased the conformity of the dose distribution to the cancerous tumor while decreasing the dose administered to the surrounding normal tissue. Even greater dose conformity has been achieved by incorporating non-coplanar beam geometries to the treatment. The non-coplanar geometry can be achieved by implementing couch rotations in one or various directions, known as pitch, roll and yaw. Pitch and roll rotations are specially achieved with the use of six-degree-of-freedom couches. However, the increased complexity of the non-coplanar treatment enhances the possibility of couch-gantry or patient-gantry collision, a safety concern. To prevent collisions from occurring, this work presents a collision avoidance computer program. It simulates a treatment plan using a linac, couch and patient model using a collision detection algorithm. Accuracy tests show a software with an average error of 2.4 cm, with some potential “blind spots” that increase the error to 4.6 cm. Data analysis suggest the need of a 3.0 cm safety buffer to increase the collision prediction capabilities of the program. This software should provide a good initial step for dosimetrists, physicists and therapists to prevent injuries and equipment damage, while improving workflow and productivity.

Degree

thesis:*
Level thesis:degree_level
Masters
Department dc:contributor.department
Mechanical Engineering
Grantor dc:publisher
Georgia Institute of Technology
Year dc:date.issued
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Sanchez Rodriguez, Jose Antonio
Advisors dc:contributor.advisor
  • Wang, C.-K. Chris
  • Schreibmann, Eduard
Committee member dc:contributor.committeemember
  • Erickson, Anna

Subjects

dc:subject × 5

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1853/59290

Chain of custody

source
Harvested from
Georgia Tech
Base URL
repository.gatech.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Sanchez Rodriguez, Jose Antonio. Experimental validation of a collision avoidance software in radiation therapy. Masters thesis, Georgia Institute of Technology, 2017. http://hdl.handle.net/1853/59290