Abstract
dc:description.abstractThis thesis considers the problems of remote temperature measurement, using a pyro-electric vidicon (PEV) to image small areas of biological tissue heated by laser irradiation. This is applied specifically to laser-assisted vascular anastomosis (LAVA), which is a surgical technique that aims to bond apposed tissue edges. It is a technique which has shown potential but the results have not been consistently successful. In order to understand the bond mechanism and to ascertain the optimum bond temperature, a number of authors have measured the temperature at the bond site using remote thermal imagers. However, if the limitations of the imager are not taken into account, then the temperature will be underestimated. These limitations were investigated extensively for the PEV but the principles apply to all thermal imagers. In particular, a minimum source width of 6 mm was required in order to avoid an underestimation of the temperature and considering the laser spot size was approximately 1.5 mm, large errors could be expected depending on the imager employed. In our case, errors of 20
Degree
thesis:*- Level dc:type.qualificationlevel
- PhD
- Grantor dc:publisher.institution
- University of Glasgow
- Year dc:date.issued
- 1995
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Gibson, Brian James
Rights
- Language dc:language
- en