dc:descriptionIN THIS STUDY THE EFFICACY OF A NEWLY DESIGNED ELECTRODE FOR AN EMERGENCY TRANSTHORACIC PACING WAS EXPERIMENTALLY EVALUATED AND THE COMPLICATIONS FROM ITS MYOCARDIAL APPLICATION WERE STUDIED. THIS NEW DEVICE CONSIST OF A THIN, FLEXIBLE INSULATED STEINLESS STEEL WIRE THAT IS INTRODUCED INTO THE RIGHT VENTRICLE USINGA 20 GAUGE THIN WALL CANULA. THE WIRE TIP REFORMS IN TWO LEVELS WHEN IT EXITS THE NEEDLE AND CONTACTS THE ENDOCARDIUM ON NEEDLE WITHDRAWAL FROM THE CHEST. THE ELECTRODE TIP DESIGN ALLOWS THE ATRAUMATIC REMOVAL OF THE WIRE WHEN PACING ISNO LONGER NEEDED. THE EXPERIMENTS WERE CARRIED OUT IN 15 MONGREL ANESTHETIZED DOGS . ALL DOGS SURVIVED THE SCHEDULED PERIOD OF ELECTRODE IMPLANTATION AND REMOVAL. NO SIGNIFICANT COMPLICATION WAS NOTED. THE MEDIAL INSERTION TIME WAS 30 SEC. A SATISFACTORY PACING ABILITY WAS ESTABLISHED THROUGH THE NEW ELECTRODE IN ALL EXPERIMENTS.