Back to search

University of Thessaly (UTH)

Monte Carlo techniques for electron radiotherapy

Abstract

dc:description

In this work simulation of electron beams and electron postmastectomy radiotherapy are evaluated using Monte Carlo Techniques. To achieve this several steps were followed. The first step was to perform measurements for dose distributions on a water tank for the different energies and field sizes provided by ELEKTA SL18 Linac available at the University Hospital of Larisa. The measurements were performed for the central axis percent depth dose (pdd), and the off-axis dose profiles at the depth of maximum dose (Dmax). Choosing, installing and using a code for the simulation of the Linac follow this. EGSnrc and the accompanying BEAMnrcMP, DOSXYZnrc and BEAMDP codes were installed. The linear accelerator was simulated using the manual provided by ELEKTA Company as a guide. Percent depth dose and dose profiles were generated and compared with the measurements and fine tuning was made to match the measurements and calculations in order to verify the simulated Linac. To reach a good match on the pdd curves several trials were performed changing the energy of the beam and the energy width, or full width at half maximum. The differences for the PDD curves were more significant in the build-up region and especially at the surface. Possible contaminating neutrons, scattered electrons from the phantom, dependence on the measuring device, the choice of the effective point of measurements and other parameters could be responsible for this mismatch. The mismatch becomes much smaller as the beam energy increases. On the other hand a slight discrepancy between measurements and calculations is noticed at the fall-off region with higher energies. But, in general these differences are within the 2%-2mm range. Regarding the off-axis dose profiles there was a good match between measurements and calculations. To get to this agreement a slight change in the jaws opening had to be made. The major discrepancy between measurements and calculations was in the right shoulder, where the measurements showed slight depression for some energies and field sizes. When one shoulder is apparently depressed and varies from the other one that is in a good match with MC calculations we consider that the profile is verified. In the second part of the work, the different sources of systematic error in linear accelerator simulations were studied. This includes slight alterations in the input beam parameters, the phantom design and placement and the configuration of the Linac components. This study aims at helping future researchers assess from the discrepancy between measured and calculated Pdd and dose profiles, where the mistake in simulation could possibly be.

Degree

thesis:*
Grantor dc:publisher
University of Thessaly (UTH)
Year dc:date
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Omer, Hiba Baha

Subjects

dc:subject × 9

Rights

Language dc:language
eng

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/18179

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Omer, Hiba Baha. Monte Carlo techniques for electron radiotherapy. University of Thessaly (UTH), 2009. http://hdl.handle.net/10442/hedi/18179