University of Ioannina
Εκτίμηση της θόλωσης του οπισθίου περιφακίου μετά από ένθεση διαφορετικών τύπων ενδοφθάλμιων φακών σε πειραματικό μοντέλο κονίκλου
Abstract
dc:descriptionPurpose: Study on cell growth on the posterior capsule after implantation of 6 different types of IOL with different design and material in a rabbit model and comparison of posterior capsule opacification (PCO). 1) Acrysof SA60AT (Alcon) made of hydrophobic acrylic. 2) Acrysof MA60BM (Alcon) made of hydrophobic acrylic. 3) Akreos Disc (Bausch & Lomb) made of hydrophylic acrylic. 4) Akreos Fit (Bausch & Lomb) made of hydrophylic acrylic. 5) CeeOn Edge 911A (Pharmacia) made of silicone. 6) Clariflex (Advanced Medical Optics) made of silicone. These IOLs had the same optic geometry (square edged) but different material and design and we wanted to study if the inhibition of PCO was influenced by different materials and designs of IOLs. Methods: Each rabbit was anesthetized. Phacoemulsification surgery through a clear corneal incision was performed in 52 rabbit eyes using an aseptic technique The IOLs mentioned above were implanted into these eyes. Ηistological sections of each globe were prepared to document the evaluation of PCO. All rabbits were evaluated at day 1 and at 1, 2 and 3 weeks for ocular inflammatory response, corneal edema and PCO, and retroillumination photos were taken with the pupil dilated. Euthanasia was performed Enucleation of the globes was performed 3 weeks after surgery. Photographs from the posterior aspect (Miyake-Apple posterior view) were performed to evaluate PCO. Central PCO (CPCO), peripheral PCO (PPCO), Sommering’s ring (SR) formation were evaluated 3 weeks after surgery using the Miyake-Apple posterior photographic technique. The intensity of CPCO, PPCO and Sommering’s ring were scored from 0 to 4. Histological sections of each globe were prepared to document the evaluation of PCO. The anterior segments of the eyes were stained with hematoxylin and eosin, periodic acid-Schiff, and Masson’s trichrome. Immunohistochemistry was performed. Type of growth, extension of PCO, cell type, inhibition, and fibrosis. The number of each cell type which was developed in different areas was estimated. Scanning Electron Microscopy study of the IOLs was also performed. Results: No statistically significant difference in central PCO or in peripheral PCO was noted between the 6 types of IOLs. No statistically significant difference in PCO or PPCO was noted among hydrophobic acrylic IOL, hydrophilic acrylic IOL and silicone IOLs. The CPCO was low in all groups. However there was statistically significant difference in Sommering’s ring (SR) being more intense SR in Acrysof SA60AT than in Acrysof MA60BM. CeeOn Edge 911A and Clariflex presented more intense SR in comparison to Acrysof MA60BM. There was a statistically significant difference between hydrophilic and non hydrophilic (hydrophobic acrylic and silicone IOLs) regarding the extension of cells, being more in the hydrophilic group. Conclusion: PCO was not influenced by the material of the IOL or the design of the haptics of the IOLs we studied.
Degree
thesis:*- Grantor dc:publisher
- University of Ioannina
- Year dc:date
- 2010
Author and committee
dc:creator, dc:contributor.*- Authors dc:creator
-
- Trakos, Nikolaos
- Τράκος, Νικόλαος
Subjects
dc:subject × 20- Θόλωση οπισθίου περιφακίου
- Υδρόφιλοι ενδοφθάλμιοι ενδοφακοί
- Ενδοφακός akreos
- Ενδοφακός acrysof sa60at
- Πειραματικό μοντέλο κονίκλου
- Ενδοφακός clariflex
- Ενδοφακός ceeon edge 911a
- Υδρόφοβοι ενδοφακοί, ενδοφακοί σιλικόνης
- Posterior capsule opacification
- Hydophylic intraocular lenses
- Akreos intraocular lens
- Acrysof intraocular lens
- Experimental rabbit model
- Clariflex intraocular lens
- Ceeon edge intraocular lens
- Hydrophobic intraocular lenses, silicone intraocular lenses
- Ιατρική και Επιστήμες Υγείας
- Κλινική Ιατρική
- Medical and Health Sciences
- Clinical Medicine
Rights
- Language dc:language
- gre
Identifiers
dc:identifier.*- Identifier
- 10.12681/eadd/22140