Εφαρμογή και αξιολόγηση του προγράμματος διασφάλισης ποιότητας στους οδοντιατρικούς υπολογιστικούς τομογράφους κωνικής δέσμης, σύμφωνα με τις ευρωπαϊκές κατευθυντήριες οδηγίες

Abstract

dc:description

Increasing numbers of CBCT units are already operative on the market. These devices use a variety of crucial exposure parameters such as the X-ray spectrum (voltage peak and filtration), X-ray exposure (mA and number of projections) and volume of the exposed field. The wide range of imaging protocols reflects to different absorbed radiation doses and overall image quality. Therefore, a Quality Assurance (QA) Protocol is needed in order to provide confidence that each CBCT unit and imaging protocol is used for its intended purpose and can be used safely in clinical application.The development of a QA protocol was one of the main aims of the SEDENTEXCT project. During this 4-year project a commercially available QA Phantom and Software were designed and developed and a QA protocol for dental CBCT units was devised. The objective measurements of the majority of imaging performance characteristics allows quantification of the factors affecting image quality, thus provide the ability to rate CBCT units in an absolute scale and monitor their performance over time. The aim of this study is the application of the QA Protocol as suggested by the SEDENTEXCT Project in CBCT units operative in Greece. For the present study the QA protocol recommended by the SEDENTEXCT Project was applied in 9 CBCT units. A total of 39 scanning protocols were used which consisted of different exposure parameters and different FOVs. For the assessment of proper functioning the piranhaRTM was used and kVp, HVL, scanning time, total filtration and radiation dose in air were recorded and compared to the ones reported by the Manufacturer of each CBCT unit. Image quality assessment of the different CBCT units and scanning protocols was made by scanning the QA Phantom which contains the test inserts for the image quality characteristics. Objective measurements were made with the use of the accompanying software for Uniformity, Noise, Artefacts, CNR (for 5 different materials: Air, Alu, PTFE, Delrin and LDPE), PSF and LSF. Subjective assessment was made for Spatial Resolution, Contrast Resolution (for 5 different materials: Air, Alu, PTFE, Delrin and LDPE) and Geometric Accuracy by 5 Oral Radiology experts, using the respective inserts included in the QA Phantom. The QA Phantom was scanned again six months later at the same CBCT units, using the same 39 scanning protocols The objective and subjective measurements were repeated and were compared to the previous ones. The suggested SEDENTEXCT QA Protocol was applicable for all CBCT units under investigation. All measurements concerning the proper functioning of the CBCT units were within the manufacturer specifications and tolerance limits. Radiation doses were varied widely (1,21-14,56mGy) between the different CBCT units and scanning protocols. Same results were recorded for HVL too (3.2-8,4mm Al), which corresponds to the vast differentiation of beam spectrum between the CBCT units. No statistical differences were found between the repeated measurements made with the Software for the same image quality parameter, implying the reliable functioning of the software. On the other hand no reliable results were recorded for the PSF and LSF and the respective measurements were rejected and not analyzed any further. There were statistically significant differences between the measurements of all image quality parameters in terms of both the CBCT units and the scanning protocols. Statistically significant differences were found also for the assessment of contrast and spatial resolution made by the observers between the CBCT units and the scanning protocols. On the contrary, measurements made for the geometric accuracy showed no statistically significant differences between the different CBCT units and scanning protocols. The QA Protocol proposed by SEDENTEXCT Project and the newly designed QA Phantom can be applicable as a routine QA process for all CBCT units investigated. A large variety of image quality parameters were measured for a wide range of CBCT units and protocols. Further work needs to be done in optimizing the phantom design for smaller FOVS and in reprogramming the accompanying Software on some measurement malfunctioning recorded. Also, more measurements must be made in order to establish the appropriate ranges for each image quality parameter, to ensure the best diagnostic value of the images produced during a CBCT scan, without exposing the patient to unnecessary risk.

Degree

thesis:*
Grantor dc:publisher
National and Kapodistrian University of Athens
Year dc:date
2014

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Makris, Nikolaos
  • Μακρής, Νικόλαος

Subjects

dc:subject × 7

Rights

Language dc:language
gre

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/31592

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Makris, Nikolaos; Μακρής, Νικόλαος. Εφαρμογή και αξιολόγηση του προγράμματος διασφάλισης ποιότητας στους οδοντιατρικούς υπολογιστικούς τομογράφους κωνικής δέσμης, σύμφωνα με τις ευρωπαϊκές κατευθυντήριες οδηγίες. National and Kapodistrian University of Athens, 2014. http://hdl.handle.net/10442/hedi/31592