Abstract

dc:description

This work characterizes experimentally the neutron spectra emitted by a high-energy medical LINAC. The measurement technique is based on neutron activation of different activation detectors. The neutron activation technique is commonly used in characterization of high neutron fluence rates present at facilities such as in nuclear reactors. This is the first attempt to apply the technique in a mixed photo-neutron field around a medical Linac. The main difficulty that arises from that implementation is that thick materials instead of thin must been used due to the much lower neutron fluence rates. The self-shielding factor is an important parameter for the calculation of the absolute nuclear reaction rates in thick materials. Monte Carlo modeling of thick materials leads to a miscalculation of the self-shielding factor due to the fact that does not take into account the neutrons scored in different energy bins (neutron leakage). A new approach based on MC simulations combined by analytical method is presented to calculate the self-shielding factor over the whole neutron energy range. The neutron self-shielding is calculated using the ratio between the reaction rates per atom in the thick and in the thin material. The thick material is the real material that were used in the experiment and the thin one represents the ideal thickness, which does not allow the cross sections of the neighboring nuclei to overlap and multiple scattering to take place. The neutron interactions within the thin material were studied using the Geant4 MC code. The neutron production and penetration through a thick material is calculated through analytical expressions developed for the thin sample. A thick material is composed of n thin slices. The neutron production from one layer to the other is calculated for each energy region using an optimized home-made routine that takes into account the possibility of a neutron to maintain its initial energy or to be scattered with lower energy. The code calculates the parameters needed to correct the neutron flux in a specific energy region due to the material's thickness. This optimized methodology improves the accuracy of the results in activation and shielding studies that takes place in many facilities.The Geant4 Monte Carlo toolkit is used to optimize the spatial arrangement of the metal foils selected for the neutron activation technique. The system of passive detectors offers the possibility to measure neutrons over a wide energy range (from thermal up to a few MeV) at intense and complex mixed n-g fields. Low-level gamma-ray spectrometry using an HPGe detector is used for the quantification of the radionuclides produced in the materials after exposure to the gamma-neutron field. Due to the variety of the geometry of the thick foils, the photo-peak efficiency for the used gamma-lines could not be estimated experimentally, so extensive simulations using the Geant4 were conducted. The experimental and the calculated photopeak efficiencies agree within 2.5% when the detector's simulation is based on both the vendors specifications and experimental optimization. This indicates that the Monte Carlo calculations of the efficiency of an HPGe detector can be used with high accuracy in experiments that incorporate targets with complicated geometry. The correction factors required for accurate γ-spectrometry measurements using thick foils are that deals with the possible loss of photons due to the self-absorption in the material and the factor that estimates the true coincidence summing (TCS) effect that is caused due to the emission of multiple gamma and X-rays in cascade with negligible time delay. To apply the self-absorption factor corrections and adjust the final photopeak efficiency, a realistic geometry of the activation detectors is built in Geant4. To calculate the coincidence summing correction factors a methodology is applied taking into account the information of the γ rays of the decayed isotope, the levels and gammas of the daughter isotope up to the max level of decayed isotope and the internal conversion coefficients.To obtain the neutron spectrum of a medical Linac from the activation technique, measurements must be unfolded. A method for unfolding the neutron energy spectra has been developed using the Minuit minimization. The unfolding procedure takes into account 15 reactions (15 activation rates) without imposing a guess solution (spectrum) a priori. The unfolding methodology has been validated with the benchmark neutron spectrum of Arkansas Nuclear One power plant. The unfolded procedure using the Minuit minimization reproduced reasonably well the physical expectation of two peaks in the spectrum. Both peaks are located in the correct energy region; from 1e-8 to 1e-6 MeV the thermal component and from 0.05 to ~1 MeV the fast component. Actually, the fast part of the spectrum gives two peaks at energies of ~ 0.1 and 0.8 MeV. The thermal peak corresponds to the 15% of the total neutron fluence and the fast component covers ~70% of the neutron spectrum.The operation of two different medical Linacs is simulated using the Geant4 to study the characteristics of 18 MeV Varian Linac and 15 MeV Elekta Linac photon beams. Simulations showed thata) The photon spectrum at the isocenter is not influenced by changes of the primary electron beam’s energy distribution and spatial spread.b) Almost all photons scored at the isocenter are primary that have only interacted with the target.c) The number of contaminant electrons is not negligible in both cases.d) The number of neutrons that are created by (γ,n) reactions at a 10x10 cm2 open field is 3.13x10-6 and 1.2x10-6 per primary electron at the isocenter for the 18 MV Varian Linac and 15 MV Elekta Linac, respectively. e) A flattening filter free beam needs less primary electrons in order to deliver the same photon fluence at the isocenter than a normal flattening filter operation.f) In the Varian Linac there is no significant increase of the surface dose due to the contaminant electrons by removing the flattening filter, whereas in the Elekta Linac there is a significant increase of the surface dose due to the contaminant electrons by removing the flattening filter due to the very thin target. To reduce the number of electrons in this case an aluminum block can be settled behind the target.g) Neutron fluence is higher at the field size of 20x20 cm2 and decreases with field size larger than of 20x20 cm2, for both Linacs.To validate the simulations results, the total neutron and photon fluence at the isocenter field were measured using the activation foil technique. The photon and neutron fluencies of the simulated experiments fall within a range of ±1 and ±2 sigma error, respectively, compared to those obtained experimentally.For an electron accelerator, the determination of bremsstrahlung photon flux is an important task. The bremsstrahlung photon flux allows us (a) to determine the characteristics of photonuclear reaction as the cross section and yield, (b) to estimate the possibility of producing the neutron-rich radioactive nuclei beam and photoneutron beam for fundamental research and (c) to determine the sensitivity of photon activation method and the possibilities for radiation protection and shielding. The measured yield from photoactivation of the targets via nine (γ,n) reactions allowed the determination of the number of photons from the threshold energy up to the end-point energy of the clinically used photon beam. The diversity of the materials allows the monitoring of the photon flux in different energy intervals. By analyzing experimental data, the integrated cross- sections for the photonuclear reaction 162Er(γ,n)161Er is calculated for 13.5 MeV bremsstrahlung end-point energy, that is unknown in the literature. To validate results the integrated photonuclear cross-section of the reaction 197Au(γ,n)196Au is calculated and compared with published data. The comparison of the cross-section data resulted in an agreement within 3%.

Degree

thesis:*
Grantor dc:publisher
Aristotle University Of Thessaloniki (AUTH)
Year dc:date
2016

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Vagena, Eleni
  • Βαγενά, Ελένη

Subjects

dc:subject × 14

Rights

Language dc:language
eng

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/37212

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Vagena, Eleni; Βαγενά, Ελένη. Study of the ionizing radiation field around accelerators. Aristotle University Of Thessaloniki (AUTH), 2016. http://hdl.handle.net/10442/hedi/37212