Back to search

National and Kapodistrian University of Athens

In vivo δοσιμετρία προσωπικού σε τμήμα PET/CT

Abstract

dc:description

Purpose: The aim of this study was to measure the radiation exposure of the staff in the first PET/CT department in Greece at the Diagnostic and Therapeutic Center of Athens HYGEIA. It is restricted for PET/CT scanner and to the use of 18F fluorodeoxyglucose (FDG). The study estimates radiation exposure to healthcare workers from patient undergoning FDG-PET imaging, toward developing more effective protective measures against radionuclide exposure. Method: To estimate the effective dose from external exposure all 7 members of the staff (2 nurses, 2 medical physicists, 2 technologists, 1 secretary) had TLD badges worn at the upper pocket of their overall, TLD rings and digital dosimeter worn at the side pocket. In addition isodose curves were measured for the distance of 20, 50,150 cm from patient who have been injected with 18F-FDG with thermoluminese detectors.Results: Dose values of PET personnel measured with digital detectors, TLD badges and TLD rings over 19 months of clinical activity consisting of a workload of about 25-28 patients /week which received 350-410MBq of 18F-FDG for a total 436 working days. Nurse1 received 2,1mSv whole body dose and 2,6 mSv hand dose, Nurse2 received 2,5mSv and 3,1mSv respectively, medical physicist1 the dose values were 2,04.mSv whole body dose and 2,97mSv hand dose and medical physicist2 received 2,01.mSv and 2,2.mSv respectively. The technologists1, 2 received 0,95 mSv whole body dose, 0,96 mSv respectively and the secretary received 0,15 mSv whole body dose.Conclusion: The personnel dose results are significantly less than the recommended annual dose by International Commission for Radiological Protection. However a greater effort should be made to reduce the doses further in line with the ALARA principple.

Degree

thesis:*
Grantor dc:publisher
National and Kapodistrian University of Athens
Year dc:date
2014

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Dalianis, Konstantinos
  • Νταλιάνης, Κωνσταντίνος

Subjects

dc:subject × 12

Rights

Language dc:language
gre

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/41074

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Dalianis, Konstantinos; Νταλιάνης, Κωνσταντίνος. In vivo δοσιμετρία προσωπικού σε τμήμα PET/CT. National and Kapodistrian University of Athens, 2014. http://hdl.handle.net/10442/hedi/41074