Εφαρμογή τρισδιάστατης χημικής δοσιμετρίας στην ακτινοθεραπεία δερματικών νόσων

Abstract

dc:description

Accurate measurements of dose profiles using kilovoltage X ray beams are challenging due to energy dependence of available dosimeters and steep dose gradients demanding high spatial resolution and precise dosimeter positioning. The scope of this thesis is to evaluate the suitability of radiochromic EBT3 films as well as the one of a novel 3D radiochromic micelle gel dosimeter, LCV-Pluronic, coupled with 3D optical computed tomography (OCT) imaging system. EBT3 films were calibrated in air for all beam qualities of a clinical unit (40-220 kVp). These data along with Monte Carlo (MC) simulations facilitated the estimation of both components of EBT3 film energy dependence in air due to variance of: the difference between dose to reference medium and dose to film (f), and the intrinsic efficiency of film mechanism producing response (kbq). Total energy dependence was found less than 3.1%. Experimental percentage depth dose (PDD) results obtained using calibration data in air showed quality and depth dependent differences, beyond film dosimetry uncertainty (<3.3), from corresponding MC simulations (up to 8%). The observed differences reduced only slightly when spectral variation as a function of depth was accounted for. Experimental data along with additional MC simulations in geometry of the irradiation set-up in water indicated variation of energy dependence in air and water, as a result of radiation field perturbation from the film oriented along the central beam axis. This implies a directional dependence of film response which necessitates that the angular distribution of photons impinging on the film is the same in the calibration and measurement geometries. Calibration of the novel radiochromic micelle dosimeter samples in cuvettes was performed based on pre- and post-irradiation spectrophotometry measurements of samples. Good linearity of dose-response and a relatively small constant value of linear fit were found (0.036±0.004). MC simulations showed excellent water equivalence of the 3D LCV-Pluronic gel dosimeter with water and corresponding calculated PDD curves were found to differ less than 1%. For the same beam quality, two different PDD experiments were performed using this dosimeter coupled with 3D OCT imaging. Data analysis revealed good agreement of background signal values between PDD experiments (0294±0.005 and 0.0313±0.008) that agree with corresponding constant values of linear fit (0.02943±0.0002, 0.0304±0.0003). After subtraction of background from response values, experimental results were found to agree with MC calculated PDD in water within experimental uncertainty (2%-5%) for both experiments (which differed <2%). Therefore, the response of the studied 3D dosimetry system seems to depend insignificantly on energy and dose-rate in contrast with other LCV-gelatin dosimeters. Although 3D dosimetry system is subjected to possible effects of temperature, ambient UV light and local imaging artifacts, it introduces a promising dosimetric tool for relative dosimetry of kilovoltage X ray beams.

Degree

thesis:*
Grantor dc:publisher
National and Kapodistrian University of Athens
Year dc:date
2020

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Prentou, Eleni
  • Πρέντου, Ελένη

Subjects

dc:subject × 33

Rights

Language dc:language
gre

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/47823

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Prentou, Eleni; Πρέντου, Ελένη. Εφαρμογή τρισδιάστατης χημικής δοσιμετρίας στην ακτινοθεραπεία δερματικών νόσων. National and Kapodistrian University of Athens, 2020. http://hdl.handle.net/10442/hedi/47823