Ανάπτυξη Monte Carlo προσομοιώσεων και δοσιμετρικών υπολογισμών σε σύγχρονα συστήματα γραμμικών επιταχυντών με ενσωματωμένο μαγνητικό τομογράφο (MR-Linacs)

Abstract

dc:description

The 1.5T/7MV Unity MR-Linac (Elekta, Crawley, UK) is gaining prominence in the field of radiotherapy. However, the presence of a strong magnetic field, which is always perpendicular to the primary photon beam, due to the Lorentz force, affects the trajectory of secondary charged particles, increasing the challenges in experimental dosimetry procedures. In addition, the absence of a dosimetry protocol or code-of-practice that accounts for the impact of the magnetic field exacerbates these challenges. Thus, this thesis aims to contribute to the existing data on MR-Linac dosimetry through Monte Carlo simulations, enabling more options during the quality assurance and End-to-end dosimetry tests in the presence of 1.5T magnetic field. The first part of this thesis presents the theoretical background of MR-Linac systems, as well as the theoretical framework of Monte Carlo simulations, which served as the primary tool for this study. The second part examines the quality assurance (QA) and dosimetry processes in 1.5T MR-Linac systems, starting with the evaluation of the computational accuracy of the EGSnrc Monte Carlo simulations. Subsequently, the dead volume of a variety of ionization chambers was determined, and its impact on computational dosimetry was assessed. As a next step, the k_(B,Q) correction factors for a 10×10cm2 reference field were calculated. The ionization chambers were positioned in four main orientations relative to the magnetic field. Corrections of up to 5% were determined for the perpendicular orientations. k_(B,Q) values were close to unity in the case of parallel orientations. An effort was made to investigate potential dependence of the k_(B,Q) correction factor in relation to the irradiation field size, irradiation depth and phantom material. In terms of relative dosimetry measurements, the beam profiles of a 10×10cm2 field were scanned along the x- and y-axis, using four ionization chambers and the microDiamond detector. Regarding the x-axis profiles, a deviation of approximately 1.55mm from the central beam axis was observed for an ideal water detector, due to the Lorentz force. Positioning the ionization chamber parallel to the magnetic field, deviations between -0.4mm up to-0.75mm were recorded after simulated y-profile scanning, attributed to the dead volume effect. These deviations were negligible when the ionization chambers were simulated parallel to the beam axis and vertical to the magnetic field. Moreover, the microDiamond detector demonstrated similar behavior to the ideal water detector. By investigating these deviations, the effective point of measurement of the detectors for profile scanning was also determined. In addition to the aforementioned active detectors, k_(B,Q) correction factors were determined for two commercially available Optically Stimulated Luminescence (OSL) dosimeters (passive detectors). The k_(B,Q) values were calculated for the three cardinal orientations (i.e., axial, sagittal and coronal orientation). Corrections of up to 6.4% were obtained, with the k_(B,Q) values being equal to unity within uncertainties in the axial orientation, regardless of the gantry’s angle of rotation. Regarding angular dependence, greater deviations were observed with the OSL dosimeters under the coronal setup. An effort was made to extend the correction factors under small fields down to 1×1cm2 for 6 small-cavity ionization chambers, 4 active solid-state detectors and a Thermoluminescent Dosimeter (TLD) microcube. Two dosimetry formalisms were considered. Using the first formalism, in which the k_(B,Q) values were determined, ionization chamber response corrections did not exceed 1.5% even for the smallest field considered. Major corrections were identified for the active solid-state detectors. Considering the second formalism, the required corrections exceeded 10%. The dependence of correction factors on phantom material was also investigated. In the last chapter of this thesis, the effect of small asymmetrical air gaps (0.1 up to 1mm) around the sensitive volume of small-cavity ionization chambers within a water-equivalent plastic phantom was studied. Ionization chamber under-response exceeded 1% even for air gaps as small as 0.1mm. Although this effect is minimized if the ionization chamber is positioned parallel to the magnetic field, it may not be considered negligible. Overall results of this thesis underscore the challenges in clinical dosimetry in the presence of a 1.5T magnetic field, and extend the relevant available data with respect to detector response. Therefore, this PhD thesis significantly contributes to the establishment of dosimetry protocols, codes-of-Practice and QA procedures in 1.5T MR-Linac systems.

Degree

thesis:*
Grantor dc:publisher
National and Kapodistrian University of Athens
Year dc:date
2025

Author and committee

dc:creator, dc:contributor.*
Authors dc:creator
  • Margaroni, Vasiliki
  • Μαργαρώνη, Βασιλική

Subjects

dc:subject × 15

Rights

Language dc:language
gre

Identifiers

dc:identifier.*
Identifier
10.12681/eadd/58279

Chain of custody

source
Harvested from
Greek National Archive of PhD Theses
Base URL
phdtheses.ekt.gr/eadd_oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Margaroni, Vasiliki; Μαργαρώνη, Βασιλική. Ανάπτυξη Monte Carlo προσομοιώσεων και δοσιμετρικών υπολογισμών σε σύγχρονα συστήματα γραμμικών επιταχυντών με ενσωματωμένο μαγνητικό τομογράφο (MR-Linacs). National and Kapodistrian University of Athens, 2025. http://hdl.handle.net/10442/hedi/58279