Abstract
dc:description.abstractWithin the scope of this thesis, a new multicolour STED microscope with a hyperspectral detection was designed and built. The challenges regarding the limitation of usable fluorophores as well as the stability of the optical set-up were overcome. Different existing spectral unmixing methods were compared to a self-implemented algorithm, both in simulations and using experimental data. A detailed study of the best separation conditions for several dye combinations was carried out for each of the methods. During the experiments, a spectral blue shift of several dyes was observed and its mechanisms were investigated. The optical resolution of the set-up was dye dependent and better than 80nm for all dyes. Moreover, an excellent separation for three fluorescent bead species and good a separation for a four-colour fixed-cell sample was achieved. This was done for the first time by purely distinguishing the different emission spectra using only one wavelength for excitation and a single STED beam for depletion.
Degree
thesis:*- Level thesis:degree_level
- thesis.doctoral
- Department thesis:degree_department
- The Faculty of Physics and Astronomy
- Grantor dc:publisher
- Universität Heidelberg
- Year dc:date.issued
- 2016
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Winter, Franziska Ruth
- Advisor dc:contributor.advisor
-
- Hell, Stefan W.
Rights
dc:rights.accessrights- Statement dc:rights.accessrights
-
- free
- Language dc:language.dcterms_iso639_2
- eng