Simultaneous measurement of ion ratios by inductively coupled plasma - mass spectrometry with a twin quadrupole instrument

Abstract

dc:description.abstract

<p>The ion beam from the inductively coupled plasma (ICP) is split into two parts by a device comprised of two electrostatic sectors in a back-to-back arrangement. Each ion beam is then monitored by a separate quadrupole mass analyzer and electron multiplier. Thus, ion ratios can be measured simultaneously without scanning. Isotope ratios for copper are measured with a precision of 0.14% relative standard deviation (RSD). Internal standard ratios for (Co+ signal)/(La+ signal) are measured with a precision of 1% RSD. Ion signals in the two channels are highly correlated, and the magnitude of flicker noise is reduced in the resulting ratios. Bias in the measured isotope ratios is substantial but can be corrected by calibration.</p>

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
dissertation
Department dc:contributor.department
Department of Chemistry
Year dc:date.issued
1996

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Warren, Arnold
Advisor dc:contributor.advisor
  • Robert Samuel Houk

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Identifier
archive/lib.dr.iastate.edu/rtd/11499/
Dc Identifier Other
12498
6455386
Mods Identifier Filename
archive/lib.dr.iastate.edu/rtd/11499/r_9635365.pdf|||Fri Jan 14 18:51:18 UTC 2022

Chain of custody

source
Harvested from
Iowa State University
Base URL
dr.lib.iastate.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Warren, Arnold. Simultaneous measurement of ion ratios by inductively coupled plasma - mass spectrometry with a twin quadrupole instrument. dissertation thesis, 1996. https://doi.org/10.31274/rtd-180813-10525