Back to search

Iowa State University

Investigation into the onset of nanoscale surface damage of biomedical grade CoCrMo: influence of contact loads, residual stresses, and environment

Abstract

dc:description.abstract

<p>"Fretting corrosion of orthopedic implants is an insidious problem and has been limiting the useful life of the implants. Fretting corrosion in artificial joints results in the formation of soluble debris that can migrate locally or systemically and may induce a cascade of inflammatory events that may ultimately result in bone loss by osteolysis and subsequent implant failure. The intent of this investigation is to examine the stress effects on dissolution parameters for the CoCrMo alloy in a physiological environment. Previous studies have shown that biomedical implants are prone to stress assisted dissolution, but the mechanics are not well understood. For this study a four point bending frame was designed to apply ""residual stress"" to the sample while the atomic force microscope (AFM) probe applies well characterized normal and shear loads across the surface. This would be analogous to a wear particle (the AFM probe) coming into contact with the implant, a failure mode commonly associated with total hip and knee prostheses. Nano-scale stress assisted dissolution experiments have been conducted on an AFM. Samples of a Cobalt Chromium (CoCr) alloy (ASTM F75), commonly used in biomedical implants, were analyzed in ambient conditions as well as in vitro physiological environment. Measured material dissolution rates are utilized to identify the critical contact loads and stresses required to initiate damage. In addition, a surface damage mechanism based on successive delamination of native oxide layer due to single asperity contact and repassivation of exposed surface is proposed to elucidate the experimental observations."</p>

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
thesis
Discipline thesis:degree_discipline
Mechanical Engineering
Department dc:contributor.department
Department of Mechanical Engineering
Year dc:date.issued
2005

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Mitchell, Andrew

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Identifier
archive/lib.dr.iastate.edu/rtd/19187/
Dc Identifier Other
20186
18125227
Mods Identifier Filename
archive/lib.dr.iastate.edu/rtd/19187/Mitchell_ISU_2005_M57.pdf|||Fri Jan 14 21:53:14 UTC 2022

Chain of custody

source
Harvested from
Iowa State University
Base URL
dr.lib.iastate.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Mitchell, Andrew. Investigation into the onset of nanoscale surface damage of biomedical grade CoCrMo: influence of contact loads, residual stresses, and environment. thesis thesis, 2005. https://doi.org/10.31274/rtd-20200616-80