New development in ICP spectrometry: low flow torch for axial viewing ICP-AES & studies in semiconductor analysis by ICP-MS

Abstract

dc:description.abstract

<p>A low argon flow torch has been developed for axially viewed ICP-AES. Outer argon flow has been drastically reduced (less than 7 L min-1 ) by cooling the torch wall with air passing through a cooling jacket added to the outside of a Fassel torch. The plasma is viewed axially through a cooled cone interface centered on the axial channel. Although detection limits (DL) for lines at ~200 nm are poorer by about a factor of two, those for lines at ~400 nm are actually much better than values typically seen for the same lines by axial viewing of a conventional, high-flow ICP.;Silicon measurement at m/z 28, 29 and 30 by ICP-MS is complicated by the spectral interferences caused by organic molecular ions, nitrogen and nitrogen oxides. A method has been developed to measure silicon in organic solvents by reducing polyatomic interferences at m/z 28 to 30 using cryogenic desolvation for removal of organic solvent from the wet droplets in aerosol, and by adding krypton or xenon to the sample gas flow of plasma. Isotope ratios can be measured by medium resolution (R = 4000) ICP-MS by direct nebulization of organic solvents. To avoid deposition on the interface cone, the flow of oxygen in the spray chamber is adjusted to titrate carbon in the plasma. Despite the sharply pointed peaks at R = 4000, ratio precision of +/-0.1% RSD can be obtained using approximately 1 mug of silicon.;Electrochemically modulated liquid chromatography (EMLC) column has been adapted online with ICP-MS to clean up blanks to reduce background signals. Metal ions can be retained from the flow of blanks in the micrometer size glassy carbon particle packing with appropriate negative potential. A Ag/AgCl (satd. NaCl) reference electrode in dilute nitric acid supporting electrolyte was used for electrochemical cell combination. Several metal ions have found to be retained with >99% extraction efficiency from the blank solution. Cleaning of the column can be done easily by stripping the deposited metal with a high positive potential.</p>

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
dissertation
Department dc:contributor.department
Department of Chemistry
Year dc:date.issued
2001

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Hasan, Towhid
Advisor dc:contributor.advisor
  • R. S. Houk

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Identifier
archive/lib.dr.iastate.edu/rtd/644/
Dc Identifier Other
1643
6078073
Mods Identifier Filename
archive/lib.dr.iastate.edu/rtd/644/r_3034188.pdf|||Sat Jan 15 01:22:42 UTC 2022

Chain of custody

source
Harvested from
Iowa State University
Base URL
dr.lib.iastate.edu/server/oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
related terms
citation

Hasan, Towhid. New development in ICP spectrometry: low flow torch for axial viewing ICP-AES & studies in semiconductor analysis by ICP-MS. dissertation thesis, 2001. https://doi.org/10.31274/rtd-180813-10882