Iowa State University
Optimization of the analytical performance of an inductively coupled plasma mass spectrometer
Abstract
dc:description.abstract<p>Methods have been found which enable a commercial inductively coupled plasma mass spectrometer to be used in several useful projects, despite its known susceptibility to ionization suppression and drifting of analyte signal. The optimization of instrument and measurement parameters for particular applications is discussed. The role of ultrasonic nebulization and internal standardization is;evaluated. Applications to uranium analysis, stable isotope tracing, flow injection analysis and liquid chromatographic detection are given; ('1)USDOE Report IS-T 1214. This work was performed under Contract W-7405-eng-82 with the U.S. Department of Energy.</p>
Degree
thesis:*- Name thesis:degree_name
- Doctor of Philosophy
- Level thesis:degree_level
- dissertation
- Department dc:contributor.department
- Department of Chemistry
- Year dc:date.issued
- 1986
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Thompson, Joseph
Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Identifier URI
- https://dr.lib.iastate.edu/handle/20.500.12876/81290
- Identifier
- archive/lib.dr.iastate.edu/rtd/8315/
- Dc Identifier Other
-
9314
6331160 - Mods Identifier Filename
- archive/lib.dr.iastate.edu/rtd/8315/r_8703774.pdf|||Sat Jan 15 02:09:54 UTC 2022