Back to search

Iowa State University

Feedback stabilized fluorescence microscope integrated with custom atomic force microscope for biological applications

Abstract

dc:description.abstract

<p>While fluorescence imaging and atomic force microscopes (AFMs) are widely used to visualize, track, and manipulate single biomolecules, the resolution of these methods is limited by sample drift. To minimize drift, active feedback methods have recently been used to stabilize single molecule microscopes on the sub-nanometer scale. However, these methods require high intensity lasers which limits their application in single molecule fluorescence measurements. Furthermore, these feedback methods cannot be used to track user-defined regions of the sample, but rather monitor the relative displacement of an unknown point on a fiducial marker. To overcome these limitations, we have developed a novel optical based approach, using low laser intensities that are compatible with single molecule fluorescence measurements, to track the displacement of a user-chosen point on a sample at high frequency and to stabilize the sample with ultra-high accuracy and precision. Our method tracks sample displacement at 8.6 kHz and stabilizes the sample with sub-nanometer resolution.</p> <p>While our ultra-stable fluorescence microscope platform can be integrated with an AFM, this requires the design of a highly customizable AFM platform. We therefore developed a custom, field programmable gate array (FPGA) controlled AFM which is readily integrated into our FPGA controlled sample scanning confocal fluorescence microscope. Our integrated AFM-fluorescence microscope is easily programmed for a variety of unique biological applications. The collimated laser design in our AFM allows for an adjustable sensitivity depending on the range of forces required, and the large range piezo (100μm in all three dimensions) presents the freedom to perform precise measurement and manipulation in a high throughput manner, over a broad range of distances. We demonstrate a high bandwidth of 1MHz with low noise levels of less than 71pN at full bandwidth and below 6pN at a relevant bandwidth of 1kHz.</p> <p>By integrating this custom AFM into our fluorescence platform, we unlock the capability to perform novel imaging and manipulation experiments. We demonstrate the ability to collect scanned fluorescence images in biological conditions while applying discrete forces to an area of interest by robustly synchronizing the sample stage scanning to lateral AFM tip motion, a brand new way to image cell dynamics under force.</p>

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
dissertation
Discipline thesis:degree_discipline
Electrical Engineering (Bioengineering)
Department dc:contributor.department
Department of Electrical and Computer Engineering
Year dc:date.issued
2020

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Schmidt, Patrick
Advisor dc:contributor.advisor
  • Sanjeevi Sivasankar

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Identifier
archive/lib.dr.iastate.edu/etd/18222/
Dc Identifier Other
9229
19236811
Mods Identifier Filename
archive/lib.dr.iastate.edu/etd/18222/Schmidt_iastate_0097E_18806.pdf|||Fri Jan 14 21:38:48 UTC 2022

Chain of custody

source
Harvested from
Iowa State University
Base URL
dr.lib.iastate.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Schmidt, Patrick. Feedback stabilized fluorescence microscope integrated with custom atomic force microscope for biological applications. dissertation thesis, 2020. https://doi.org/10.31274/etd-20200902-141