Cadmium Zinc Telluride (CdZnTe) has been used for many applications, such as medical imaging and astrophysics, since its first demonstration as a room temperature operating gamma-ray detector in 1992. The wide band gap, high effective Z-number and high resistivity of CdZnTe make it a good candidate for use as a room temperature operated detector with good absorption efficiency, while maintaining a low bulk leakage current at high electric fields. Nevertheless, the low mobility lifetime products mu tau of holes in CdZnTe makes detectors position sensitive, unless advanced detector designs are employed. Among those designs is the Frisch collar technology which turns the detector into a single carrier device by negating the degrading effects of hole trapping and low mobility. The superiority of the Frisch collar technology over other methods include its inexpensive associated electronics and straight forward fabrication process. The main objective of this research study is to develop a large volume gamma-ray detector with an array of individual CdZnTe Frisch collar gamma-ray spectrometers while still using a single readout. Several goals were to be accomplished prior to the main objective. One goal is to develop a reliable low cost method to fabricate bulk CdZnTe crystals into Frisch collar detectors. Another goal was to investigate the limitations of crystal geometry and the crystal electrical properties to obtain the best spectroscopic performance from CdZnTe Frisch collar detectors. Still another goal was to study all other external parameters such as the collar length, anode to cathode ratio, the insulator thickness and applied voltage on performance of CdZnTe Frisch collar detectors. The final goal was to construct the CdZnTe Frisch collar devices into an array and to show its feasibility of being used for large volume detector.

Degree

thesis:*
Level thesis:degree_level
Doctoral
Department dc:contributor.department
Department of Mechanical and Nuclear Engineering
Grantor dc:publisher
Kansas State University
Year dc:date.issued
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Kargar, Alireza
Advisor dc:description.advisor
  • Douglas S. McGregor

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • © the author. This Item is protected by copyright and/or related rights. You are free to use this Item in any way that is permitted by the copyright and related rights legislation that applies to your use. For other uses you need to obtain permission from the rights-holder(s).
Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/2097/1665

Chain of custody

source
Harvested from
Kansas State University
Base URL
krex.k-state.edu/server/oai/request
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Kargar, Alireza. Characterization and optimization of CdZnTe Frisch collar gamma-ray spectrometers and their development in an array of detectors. Doctoral thesis, Kansas State University, 2009. https://hdl.handle.net/2097/1665