Back to search

Kansas State University

Rhinitis radiofrequency ablation: FEM analysis and experiments

Abstract

dc:description.abstract

The primary objective of this research is to implement an experimentally validated computational model to guide device design and selection of energy delivery strategies for treating chronic rhinitis by radiofrequency ablation. Chronic rhinitis is one of the most common global health problems. It is not life-threatening but has a severe impact on quality of life. Direct cost by chronic rhinitis is enormous and places a burden on societies. Radiofrequency ablation is proposed to be as an efficient treatment providing symptom relief and avoiding side effects compared to traditional therapies. Three-dimensional finite element method (FEM) models were developed to investigate RFA devices and energy delivery strategies. FEM computational models could provide vital variable profiles that are technically challenging to determine through experiments. Also, computer simulation could reduce the number of experimental procedures during the device design process. First, single pair bipolar RF ablation experiments were performed to validate FEM simulations using the same geometry as in experiments. The data from experiments and simulations had a high correlation (R = 0.91). Second, the Neurent basket electrode was employed in experimental ablations in egg white, for comparison against FEM simulations. Smaller lesion sizes were observed in experiments compared to simulations, attributed to thermal convection that was not accounted for in simulation. Finally, FEM simulations were used to investigate the effects of basket electrode diameter, length, and applied power on ablation zone formation. A shorter but wider electrode with a maximum spacing distance between two pairs of electrodes is preferable to create discontiguous ablation zones. 50% duty cycle was recommended to create thermal ablation zones with gradually increasing temperature and sufficiently large thermal lesion volumes.

Degree

thesis:*
Level thesis:degree_level
Masters
Department dc:contributor.department
Department of Electrical and Computer Engineering
Grantor dc:publisher
Kansas State University
Year dc:date.issued
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Song, Yuqi
Advisor dc:description.advisor
  • Punit Prakash

Subjects

dc:subject × 4

Rights

dc:rights
Statement dc:rights
  • © the author. This Item is protected by copyright and/or related rights. You are free to use this Item in any way that is permitted by the copyright and related rights legislation that applies to your use. For other uses you need to obtain permission from the rights-holder(s).
Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/2097/39132

Chain of custody

source
Harvested from
Kansas State University
Base URL
krex.k-state.edu/server/oai/request
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Song, Yuqi. Rhinitis radiofrequency ablation: FEM analysis and experiments. Masters thesis, Kansas State University, 2018. https://hdl.handle.net/2097/39132