Back to search

University of Manitoba

Sensitivity in MALDI MS with small spot sizes

Abstract

dc:description.abstract

In MALDI, for laser fluences below the saturation point the ion yield per shot follows a cubic dependence on the irradiated area, leading to a conclusion that smaller spots produce overall less ions and therefore are less viable. However, Qiao et al. showed that by decreasing the laser spot size it is possible to raise the saturation point, and thus increase the ion yield per unit area, also known as sensitivity. Here we explore laser spots below 10 micrometer diameter to determine whether they offer any practical advantage. We show that sensitivity is greater for a flat-top 3-4 micrometer spot than for a 10 micrometer spot. The sensitivity is greater for a Gaussian-like 3-5 micrometer spot than for flat-top 5-25 micrometer spots. We also report for the first time sensitivity versus theoretical fluence profile for a Gaussian-like beam focused to a spot of 3-5 micrometer.

Degree

thesis:*
Level thesis:degree_level
Master of Science (M.Sc.)
Discipline thesis:degree_discipline
Physics and Astronomy
Year dc:date.issued
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Yamchuk, Andriy
Advisor dc:contributor.supervisor
  • Ens, Werner (Physics & Astronomy)
Committee members dc:contributor.committeemember
  • Standing, Ken (Physics & Astronomy)
  • Krokhin, Oleg (Chemistry)

Subjects

dc:subject × 9

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1993/23227

Chain of custody

source
Harvested from
University of Manitoba
Base URL
mspace.lib.umanitoba.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Yamchuk, Andriy. Sensitivity in MALDI MS with small spot sizes. Master of Science (M.Sc.) thesis, 2014. http://hdl.handle.net/1993/23227