in vivo patient dose verification of volumetric modulated arc therapy including stereotactic body radiation treatment applications using portal dose images
Abstract
dc:description.abstractThe complexity of radiation therapy delivery has increased over the years due to advancements in computing and technical innovation. A system of dose delivery verification has the potential to catch treatment errors and therefore improve patient safety. The goal of this thesis was to create a portal image-based in vivo dose reconstruction model for volumetric modulated arc therapy (VMAT) deliveries, specifically for stereotactic body radiation therapy (SBRT). This model-based approach should be robust and feasible within a clinical setting. VMAT involves the modulation of dose rate, gantry speed, and aperture shaping while the treatment gantry (i.e., x-ray beam) rotates about the patient. In this work, portal images were acquired using an amorphous silicon electronic portal imaging device (a-Si EPID). A geometrical characterization of the linear accelerator (linac) during VMAT delivery was performed. An angle adjustment method was determined which improves each EPID’s angular accuracy to within ±1° of the true physical angle. SBRT delivers large doses over fewer fractions than conventional radiotherapy, therefore, any error during an SBRT delivery will have a greater impact on the patient. In this work, a robust, model-based SBRT-VMAT dose reconstruction verification system using EPID images was developed. The model was determined to be clinically feasible. The accuracy of a 3D in vivo dose reconstruction, using all the EPID images acquired during treatment, is sensitive to the chosen frame averaging per EPID image: the greater the frame averaging, the larger the reconstruction error. Optimization of the EPID frame averaging number as a function of average linac gantry speed and dose per fraction were determined. The EPID-based in vivo dose reconstruction model for SBRT-VMAT developed here was determined to be robust, accurate, and clinically feasible as long as adjustments were made in order to correct for EPID image geometrical errors and frame-averaging errors.
Degree
thesis:*- Level thesis:degree_level
- Doctor of Philosophy (Ph.D.)
- Discipline thesis:degree_discipline
- Physics and Astronomy
- Grantor dc:publisher
- Medical Physics
- Year dc:date.issued
- 2015
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- McCowan, Peter Michael
- Advisor dc:contributor.supervisor
-
- McCurdy, Boyd (Physics & Astronomy)
- Committee members dc:contributor.committeemember
-
- Rickey, Daniel (Physics and Astronomy)
- Lewis, John (Physics and Astronomy)
- Ryner, Lawrence (Physics & Astronomy)
- Lin, Francis (Physics and Astronomy)
- Thomas, Gabriel (Electrical and Computer Engineering)
Subjects
dc:subject × 2Rights
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Dc Identifier Citation
- MLA