Development of microneedle array electrodes for transcutaneous neural stimulation and recording applications

Abstract

dc:description.abstract

Transcutaneous electrical nerve stimulation is one of the physical neuromodulation methods that stimulate a nerve and potentially its adjacent nerve fibers. To make this method more effective, it would be of value to design new electrodes that only target a specific nerve and prevent the co-stimulation of an adjacent nerve. In this dissertation, different scenarios of tissue and blood capillary placements were analyzed through multiphysics simulation of a 3D model and the best microneedle array electrode was suggested as a new electrode for nerve stimulation. These studies showed that the design of the microneedle array plays a critical role in presenting a uniform electric field to a target nerve and the resulting performance benefits. Analysis showed that microneedle array electrodes can increase the distinguishability between a target nerve fiber and its adjacent nerve fibers up to 10% compared to conventional surface electrodes. Uniform electric field, effective thermal behaviour and consistent current density in different depths of tissue are the other advantages of using microneedle electrodes for stimulation. Different stimulation parameters were investigated and the analysis showed that the geometry of the microneedles on top of the electrodes, such as their tip-to-tip pitch, their numbers and shape can affect performance. Microneedle arrays were fabricated and used for experimental verification of performance benefits. One study explored polymer microneedles molded from a polycarbonate master. They were coated with titanium and titanium-nitride to increase the mechanical strength, and molybdenum (which is a biocompatible metal) to increase the conductivity of the structure. In clinical study it was found that they enhanced the tissue-electrode interface parameters, such as decreasing the resistance and increasing the double-layer capacitance. In another study, low cost microneedle arrays were printed using 3D printing technology. In a clinical study comparing with the molded microneedles, it was found that the 3D printed ones had better high frequency performance (such as for EMG), and the molded microneedles performed better for low frequency signals (such as for EEG).

Degree

thesis:*
Level thesis:degree_level
Doctor of Philosophy (Ph.D.)
Discipline thesis:degree_discipline
Biomedical Engineering
Year dc:date.issued
2019

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Soltanzadeh, Ramin
Advisor dc:contributor.supervisor
  • Cyrus, Shafai (Electrical and Computer Engineering)
Committee members dc:contributor.committeemember
  • Sherif, Sherif (Electrical and Computer Engineering)
  • Yahampath, Pradeepa (Electrical and Computer Engineering)
  • Blakley, Brian (Otolaryngology)
  • Casson, Alex (School of Electrical and Electronic Engineering, The University of Manchester)

Subjects

dc:subject × 3

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Dc Identifier Citation
Soltanzadeh, Ramin, et al. "Molybdenum coated SU-8 microneedle electrodes for transcutaneous electrical nerve stimulation." Biomedical microdevices 20.1 (2018): 1.

Chain of custody

source
Harvested from
University of Manitoba
Base URL
mspace.lib.umanitoba.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Soltanzadeh, Ramin. Development of microneedle array electrodes for transcutaneous neural stimulation and recording applications. Doctor of Philosophy (Ph.D.) thesis, 2019. https://hdl.handle.net/1993/34297