Digital Repository at the University of Maryland
Signal Corrections and Calibrations in the LUX Dark Matter Detector
Abstract
dc:description.abstractThe Large Underground Xenon (LUX) Detector has recently finished a 332-day exposure and placed world-leading limits on the spin-independent WIMP-nucleon scattering cross section. In this work, we discuss the basic techniques to produce signal corrections, energy scale calibrations, and recoil band calibrations in a dark matter detector. We discuss a nonuniform electric field that was present during LUX's 332-day exposure, and detail how such a field complicates these calibration techniques. Finally, we expand on the techniques presented earlier, such that they account for all of the complication introduced by the nonuniform electric field, and allow a limit to be produced from the data.
Degree
thesis:*- Department dc:contributor.department
- Physics
- Grantor dc:publisher
- Digital Repository at the University of Maryland
- Year dc:date.issued
- 2016
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Knoche, Richard
- Advisor dc:contributor.advisor
-
- Hall, Carter
Rights
- Language dc:language.iso
- en
Identifiers
dc:identifier.*- Identifier URI
- http://hdl.handle.net/1903/19300