RESPIRATORY MECHANICS OF FLOW LIMITATION AND CHARACTERIZATION OF RESISTANCE MEASUREMENTS WITH A NON-INVASIVE DEVICE

Abstract

dc:description.abstract

Resistance measurements with the airflow perturbation device (APD) were compared to directly measured pulmonary resistances with an esophageal balloon to validate the APD. The APD perturbs the flow and the mouth pressure during regular breathing. The ratio of mouth pressure perturbations to the flow perturbations was used to calculate the inspiratory, expiratory and average respiratory resistance. Six healthy subjects were tested during tidal breathing when known external resistances were added during inspiration, during expiration, and during both inspiration and expiration. The difference between the averaged APD measured and directly measured pulmonary resistances was 0.59 &plusmn 1.25 (mean &plusmn SD) cmH<sub>2</sub>O/L/s. Compared to the magnitude of the known increase in added resistance, the APD measured resistance increased by 79 %, while the directly measured pulmonary resistance increased only by 56%. During addition of external resistances to both inspiration and expiration, the changes in inspiratory and expiratory pulmonary resistance were only 36 % and 62 % of the added resistance, respectively. On the other hand, the APD inhalation and exhalation resistance measured between 82 % and 76 % of added resistance change. It was concluded that the APD detects changes in external resistance at least as well and probably better than classical measurements of pulmonary resistance. Additionally, expiratory isovolume pressure - flow (IVPF) curves, which show the pressure at which the flow becomes limited during forced expiration, were constructed in six healthy subjects with the classical invasive method of esophageal balloon (EB) and the alternative noninvasive method of stop - flow (SF) at 25, 50, and 75 % vital capacity (VC). The difference between the pressures (P<sub>max</sub>) and flow (Q<sub>max</sub>) at which flow limitation first occurs and correlation with the stop - flow and esophageal balloon methods were studied. Additionally, the resistance at flow limitation was compared to the APD resistance during forced breathing. On average, P<sub>SF,max </sub>was 5.6 and 4.4 times P<sub>EB,max</sub> at 25 %VC and 50 %VC, respectively. Q<sub>SF,max </sub>was 0.68 and 0.59 times Q<sub>EB,max </sub>at 25 %VC and 50 %VC, respectively. No correlation was found between the stop - flow and esophageal balloon methods as well as between the resistances at flow limitation.

Degree

thesis:*
Department dc:contributor.department
Biological Resources Engineering
Grantor dc:publisher
Digital Repository at the University of Maryland
Year dc:date.issued
2009

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Coursey, Derya
Advisor dc:contributor.advisor
  • Johnson, Arthur T

Rights

Language dc:language.iso
en_US

Identifiers

dc:identifier.*
Handle dc:identifier.uri
http://hdl.handle.net/1903/9272

Chain of custody

source
Harvested from
University of Maryland
Base URL
api.drum.lib.umd.edu/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Coursey, Derya. RESPIRATORY MECHANICS OF FLOW LIMITATION AND CHARACTERIZATION OF RESISTANCE MEASUREMENTS WITH A NON-INVASIVE DEVICE. Digital Repository at the University of Maryland, 2009. http://hdl.handle.net/1903/9272