Development of a Microchip-Based Flow Cytometer with Integrated Optics – Device Design, Fabrication, and Testing

Abstract

dc:description.abstract

<p>Lab-on-a-chip technologies have created a burgeoning number of new and novel devices designed to automate biological processes on-chip in an efficient and inexpensive format for far reaching point-of-care (POC) medicine and diagnostic treatments and for remote and on-line monitoring functions. This work designed a device that integrated advanced optical functionality on-chip with the microfluidics to relieve the reliance on traditional bulky and expensive free-space optics and a high-quality light source. The multimodal input beam was reshaped into an optimized geometry in the microchannel via a 2D system of lenses - improving the quality and reliability of detection through uniform detection of particles. A uniform beam geometry across the sample stream with a uniform beam width will allow repeatable excitation and burst duration to allow for more reliable and predictable detection. Numerous beam geometries were created and the quality and illumination properties confirmed by testing each with a couple sizes of fluorescent and non-fluorescent microspheres to test the effect of beam geometry and particle size combination on device performance. The measured coefficient of variation (CV) for fluorescent beads was found to have a particular beam geometry that yielded best device performance based on the bead size. Fluorescent beads 2.5µm in diameter had a CV of 8.5% for a 3.6 µm beam waist while 6 µm beads yielded a 14.6% CV with a 10 µm beam waist. When measuring scatter and fluorescence signal from a 10 µm the 2.5- and 6.0 µm beads gave 11.4% and 15.8% and 15.9% and 20.4% fluorescent and scatter CVs for each set of beads, respectively. Separately testing each beam geometry with 1-, 2-, and 5 µm beads did not yield any predictable ideal beam-bead ideal pairing for best performance. Lastly, further integration of optical function was shown through the on-chip collection of signals; CVs of 29% and 30% were measured for side scatter and forward scatter, respectively, for 5 µm beads. The reliability of this all-optically guided scheme was confirmed by comparing it to a simultaneously recorded free-space collection scheme. The coincidence rate was found to be 94% and 96% for the side scatter and forward scatter schemes. Both had very low false positive rates – below 0.5% - with missed detection rates that were satisfactory but in need of improvement. Sources of noise and device improvements were identified and suggested.</p>

Degree

thesis:*
Department dc:contributor.department
Engineering Physics
Year dc:date.issued
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Watts, Benjamin
Advisors dc:contributor.advisor
  • Chang-Qing Xu/Zhiyi Zhang
  • Fang, Qiyin
  • Selvaganapathy, Ravi

Subjects

dc:subject × 6

Identifiers

dc:identifier.*
Dc Identifier Other
opendissertations/7855
8840
4061825

Chain of custody

source
Harvested from
McMaster University
Base URL
macsphere.mcmaster.ca/server/oai/request
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Watts, Benjamin. Development of a Microchip-Based Flow Cytometer with Integrated Optics – Device Design, Fabrication, and Testing. 2013. http://hdl.handle.net/11375/13020