McMaster University
Coincidence Counting Techniques For The Determination of ²²⁶Ra Concentrations in Water
Abstract
dc:description.abstract<p>Three coincidence counting techniques for the assay of ²²⁶Ra in water samples are evaluated and compared. The techniques are based on the detection of alpha-gamma, beta-gamma, and gamma-gamma coincident radiations. Samples were prepared by co-precipitating ²²⁶Ra with barium sulphate. Details of this chemical procedure are presented in this thesis. The lower limit of detection, sensitivity to sample thickness, and system precision are investigated. Finally, two data analysis techniques, the method of Least Squares and the method of Maximum Likelihood, for determining -ray peak areas as applied to low counting rates, are discussed.</p>
Degree
thesis:*- Department dc:contributor.department
- Physics
- Year dc:date.issued
- 1981
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Lowe, Diane M.
- Advisor dc:contributor.advisor
-
- Prestwich, W. V.
Subjects
dc:subject × 1Identifiers
dc:identifier.*- Identifier URI
- https://doi.org/10.71548/14850
- Dc Identifier Other
-
opendissertations/3632
4649
1671460