University of Missouri--Columbia
Finite element analysis for the insertion process of the press-fitted total hip arthroplasty
Abstract
dc:description.abstract[ACCESS RESTRICTED TO THE UNIVERSITY OF MISSOURI-COLUMBIA AT AUTHOR'S REQUEST.] There are more than 300,000 total hip arthroplasty (THA) surgeries performed every year in the United States. Press-fitted artificial hip joints are wildly used for relatively young patients. It relies on the friction resistance to relative motion at the bone-implant interface to allow bone ingrowth and primary stability. Interference fit and friction coefficient are two important factors during the insertion process that can directly influence the stability of the implants. The three-dimensional dynamic model was evaluated using finite element method to simulate the insertion process of the stem prosthesis. Different values of interference fit (from 0.1 to 0.5 mm) and friction coefficients (from 0.3 to 0.5) were considered.
Degree
thesis:*- Name thesis:degree_name
- M.S.
- Level thesis:degree_level
- Masters
- Discipline thesis:degree_discipline
- Chemistry (MU)
- Grantor dc:publisher
- University of Missouri--Columbia
- Year dc:date.issued
- 2018
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Zhao, Yanning
- Advisor dc:contributor.advisor
-
- El-Gizawy, A. Sherif (Ahmed Sherif), 1945-
Rights
dc:rights- Statement dc:rights
-
- Access to files is limited to the University of Missouri--Columbia with SSO login.
- Language dc:language.iso
- eng, English
Identifiers
dc:identifier.*- Identifier URI
- https://hdl.handle.net/10355/66208