Software tool for performing the calibration of the AT-TPC electronics channels

Abstract

dc:description

"The Active Target Time Projection Chamber (AT-TPC), developed at the National Superconducting CyclotronLaboratory (NSCL) at Michigan State University, is a gas-based system that serves a dual role of a target and detector medium for nuclear physics experiments. Active target devices have gained much attention within the past dedicate due to their high resolution and efficiency to study reactions with very exotic nuclei. With the use of incident beam energy ranging from low (1-10 MeV/u) to medium (10-100 MeV/u) to extract the physical properties of nuclei far from stability, the AT-TPC is contributing significantly to experimental nuclear science, thanks to its high luminosity working capability without loss of resolution and low-energy detection thresholds that enable experiments with beam intensities as low as 102 pps. Because of its unique features, a dedicated python based data analysis tool was developed to extract information from this detector. This software tool allows to calibrate the electronic baselines, amplitudes and times of the signals recorded by each of the AT-TPC pads relative to each other. The statistical analysis of the aforementioned calibrated parameters shows the coefficient of the fluctuations of the order of 9.1 % and 8.8 % for the baselines , 3.43 % and 4.8 % for the amplitudes and 45 ns and 34 ns for the time from the big and small pads respectively from all 24 events taken during the pulse run."--Page ii.

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Ndayisabye, Felix
Contributors dc:contributor
  • Bazin, Daniel
  • Zegers, Remco
  • Spyrou, Artemis

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • Attribution-NonCommercial-ShareAlike 4.0 International
Language dc:language
English

Identifiers

dc:identifier.*
Identifier
etd:47816
isbn:9781392187296
isbn:139218729X
oclc:1117774666
umi:13865668
local:NDAYISABYE_grad.msu_0128N_16758

Chain of custody

source
Harvested from
Michigan State University
Base URL
d.lib.msu.edu/oai
Last updated
2026-11-18
Source record
OAI-PMH GetRecord
citation

Ndayisabye, Felix. Software tool for performing the calibration of the AT-TPC electronics channels. 2019. https://doi.org/doi:10.25335/5s31-ad37