Back to search

NJIT

Late deflation study : hemodynamic effects of IAB timing in humans

Abstract

dc:description.abstract

Conventional IABP timing deflates 100% of the balloon volume before the onset of left ventricular ejection and has been well established in the literature as a safe and effective method. Yet, recent studies suggest that deflation of the IAB at or near the onset of systole improve myocardial efficiency (Kern, 1999, 1129). To test whether deflation of the JAB at a later deflation time reduces left ventricular workload and enhances coronary perfusion, four timing methods were evaluated in 20 patients: T_1 (100% JAB volume deflated before onset of ejection), T_2 (60% IAB volume deflated before and 40% volume during ejection), T_3 (25% IAB volume deflated before and 75% volume during ejection) and T_4 (100% deflation simultaneous with left ventricular ejection). To identify an optimal timing point for the deflation of the JAB, data was analyzed. Data analysis consisted of three main parts: elimination of files containing noise artifact, normalization of data for ease of analysis, and correction for variations in mean aortic pressure and heart rate which take into account the dynamic state of the heart. Late LAB deflation at 50% of the volume deflated prior to left ventricular ejection produced significantly greater percentage changes in systolic pressure time index (SPTI), diastolic pressure time index (DPTI), the SPTI/DPTI ratio, systemic vascular resistance (SVR) as compared to conventional timing, T_1. No significant change in heart rate or cardiac output was established.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Engineering - (M.S.)
Discipline thesis:degree_discipline
Biomedical Engineering Committee

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Morcos, Manal
Contributors dc:contributor
  • R. S. Sodhi
  • David S. Kristol
  • John Tavantzis

Subjects

dc:subject × 4

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.njit.edu/theses/781

Chain of custody

source
Harvested from
NJIT
Base URL
digitalcommons.njit.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Morcos, Manal. Late deflation study : hemodynamic effects of IAB timing in humans. https://digitalcommons.njit.edu/theses/781