NJIT
A study of platinum, palladium, bismuth, copper, and silver for the potentiometric determination of chlorine
Abstract
dc:description.abstractThis research was undertaken to determine which metals would be suitable to be used as electrodes for the determination of the concentration of chlorine in aqueous solutions at various pH levels. Platinum, palladium, bismuth, copper and silver electrodes were constructed and tested to determine their applicability for measuring the concentration of chlorine in drinking water, waste water treatment plant effluent, and swimming pools. Buffer solutions were used which varied in pH from 7 to 8, and in cHorine concentration from 0 to 10 ppm. In addition, the potential variation over time was observed for each electrode in the solutions. Platinum, palladium, and copper were found the most appropriate metals for the potentiometric determination of chlorine. The data obtained using these electrodes conformed with Nernstian theory. In addition, the response for palladium appeared to be affected by a variation of pH.
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Environmental Science - (M.S.)
- Discipline thesis:degree_discipline
- Chemical Engineering, Chemistry and Environmental Science
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Gottwald, Maria A.
- Contributors dc:contributor
-
- Barbara B. Kebbekus
- Richard B. Trattner
- Kenneth Rudolph Farmer
Subjects
dc:subject × 4Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.njit.edu/theses/794