Development of a multi-lead trans-thoracic impedance pneumograph for heart rate variability studies

Abstract

dc:description.abstract

In this study, a device that measures respiratory tidal volume and ECG for use in heart rate variability studies, was designed, built and tested. The device developed is a Multi-Lead Trans-Thoracic Impedance Pneumograph, which determines tidal volume on a breath by breath basis, utilizing both hardware and software. It uses standard ECG electrodes which provide increased comfort and inability to the subject over traditional spirometry methods. The device operates similar to traditional bipolar impedance pneumographs except that instead of only one lead (electrode pair), four leads are acquired simultaneously using time-division multiplexing; providing data from different locations and vectors on the subjects chest. Respiration, ECG and spirometry data were collected using the device from 6 normal, healthy subjects for several trials of normal and paced breathing. Results were correlated against the spirometry data to determine whether sufficient data are provided by the four leads and also to determine the accuracy of the device. It was concluded that combining data from all 4 leads using generalized additive modeling results in correlations in the range of 0.8 - 1.0, but more investigation is needed to determine usefulness in actual HRV applications.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Engineering - (M.S.)
Discipline thesis:degree_discipline
Biomedical Engineering Committee

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Bashford, Curtis
Contributors dc:contributor
  • Stanley S. Reisman
  • David S. Kristol
  • Peter Engler

Subjects

dc:subject × 5

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.njit.edu/theses/1144

Chain of custody

source
Harvested from
NJIT
Base URL
digitalcommons.njit.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Bashford, Curtis. Development of a multi-lead trans-thoracic impedance pneumograph for heart rate variability studies. https://digitalcommons.njit.edu/theses/1144