Abstract

dc:description.abstract

Osteosarcoma is the most common malignant bone tumor in the pediatric age group. This thesis presents the analysis of a total femoral prosthesis implanted in a 9 year old black male. The design includes a bipolar, expanding mid-shaft component, and a rotating hinge knee used in the reconstruction of the patient's right lower limb after surgical resection of the femur. The areas being analyzed are (i) the trochanteric attachment, (ii) the expanding component, and (iii) the rotating hinge knee. The stresses acting on the cables reconnecting the greater and lesser trochanter were determined for bipedal stance, single-legged stance, stance phase of gait, running and they were calculated to be 0.5 MPa, 17 MPa, 23 MPa and 51 MPa respectively. On comparing these stresses to the fatigue data for stainless steel cables, it is estimated that the cables will last for 10^8 cycles The expanding component cross-pin was analyzed to experience loads of up to three times the body weight. The maximum stress required for failure is 275 MPa at 10^7 cycles. The cross-pin is not expected to fail in fatigue. The cylindrical bearing surfaces in the rotating hinge knee experience contact stresses of 11 MPa, which is higher than the maximum stress under fatigue loading of polyethylene. The polyethylene bearing component will need to be replaced at about 10^4 cycles. Rotating knee hinge-pin component was analyzed in bending and shear, the low stresses (100 MPa and 11 MPa respectively) on the pin would give a working life of more than 10^7 cycles. The carriage-pin experiences a maximum bending stress of 98.16 MPa which is less than 50% of the maximum stress required for the titanium alloy to fail at 10^7 cycles.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Engineering - (M.S.)
Discipline thesis:degree_discipline
Biomedical Engineering Committee

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Gundlapalli, RamaRao Gundlapalli, V.
Contributors dc:contributor
  • Clarence W. Mayott
  • Mark Zimmerman
  • J. Russell Parsons

Subjects

dc:subject × 3

Identifiers

dc:identifier.*
Repository record dc:identifier
https://digitalcommons.njit.edu/theses/1759

Chain of custody

source
Harvested from
NJIT
Base URL
digitalcommons.njit.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Gundlapalli, RamaRao Gundlapalli, V.. An analysis of a custom expanding total femoral prosthesis with a rotating hinge knee. https://digitalcommons.njit.edu/theses/1759