Abstract

dc:description.abstract

<p>A new method of measuring range for gait analysis is described. The method is based on a coherent measurement of the phase shift of an ultrasonic signal as a function of range.</p><p>A prototype of the ranging system was breadboarded, tested, and found to have a relative accuracy of better than 0.5 mm over a range of 3 meters. The maximum range the system is at least 20 meters. The method used here is new and has not been found in the literature. Tt is demonstrated by experiments with a prototype system that can measure distance with a higher accuracy than the traditional pulse-echo method at a acoustic frequency of 40 KHz. A 3-dimensional system using one ultrasonic transmitter and at least four receivers to perform a 3-D ranging is discussed. Tt can be applied to many fields, including gait analysis.</p>

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Engineering - (M.S.)
Discipline thesis:degree_discipline
Biomedical Engineering Committee

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Wu, Mu
Contributors dc:contributor
  • Peter Engler
  • Bernard Friedland
  • Eugene I. Gordon