Abstract
dc:description.abstract<p>One important aspect for reconstructing an image using Computed Tomography (CT) is that the axis-of-rotation must pivot at the same point on the reconstruction plane as the x-ray source and CT scanner assembly rotates around the imaged object. This pivot point is used as a reference point for backprojecting pixel values to their proper coordinates. Therefore, reconstructing an image with a non-stationary axis-of-rotation would backproject pixel values to wrong coordinate points, since the reference point is non-stationary for every projection angle.</p><p>A convolution filtered backprojection algorithm has been derived for correcting images that were acquired with a non-stationary axis-of-rotation using the fan beam geometry with a collinear (flat) detector. Fan beam reconstruction algorithms which filter and backproject the projection data without correcting for this misalignment would produce structured artifacts and sharp degradation of the resulting image.</p><p>The correction method described in this thesis accounts only for the vertical displacements of the axis-of-rotation as the CT scanner rotates around the imaged object as may be the case when sagging occurs.</p><p>Software simulations are performed to show how the algorithm corrects for the shift in the axis-of-rotation.</p>
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Electrical Engineering - (M.S.)
- Discipline thesis:degree_discipline
- Electrical Engineering
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Concepcion, Joseph A.
- Contributors dc:contributor
-
- G. Kuo-Petravic
- John D. Carpinelli
- Stanley S. Reisman
Subjects
dc:subject × 3Identifiers
dc:identifier.*- Repository record dc:identifier
- https://digitalcommons.njit.edu/theses/2469