Monterey, California ; Naval Postgraduate School
Design and analysis of a resistively heated surgical probe
Abstract
dc:description.abstractThis thesis examines the use of heat for producing clinical lesions in tissue and presents the design and analysis of a resistively heated surgical probe. The probe temperature is accurately maintained and measured by using a Wheatstone bridge. The probe was embedded in an agar-water test medium and the temperature field emanating from the probe was measured using liquid crystal tapes. The experimental results compare within approximately 10% of a two-dimensional numerical solution. A one-dimensional theoretical model is developed and results are presented to show the relative effects of metabolism and blood flow on the temperature field.
Degree
thesis:*- Name thesis:degree_name
- M.S. in Mechanical Engineering
- Level thesis:degree_level
- Masters
- Discipline thesis:degree_discipline
- Mechanical Engineering
- Department dc:contributor.department
- Mechanical Engineering
- Grantor dc:publisher
- Monterey, California ; Naval Postgraduate School
- Year dc:date.issued
- 1971
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Groff, Jon Philip
- Advisor dc:contributor.advisor
-
- Cooper, Thomas E.
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- This publication is a work of the U.S. Government as defined in Title 17, United States Code, Section 101. Copyright protection is not available for this work in the United States.
- Approved for public release; distribution is unlimited.
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/10945/15645