Development of an automated scanning monochromator for sensitivity calibration of the MUSTANG instrument.

Abstract

dc:description.abstract

An Automated Scanning Monochromator was developed using an Acton Research Corporation (ARC) monochromator, Ealing photomultiplier tube and a Macintosh PC, in conjunction with LabVIEW software. The LabVIEW Virtual Instrument written to operate the ARC monochromater is a mouse-driven user-friendly program developed specifically for automated spectral data measurements. Resolution and sensitivity of the Automated Scanning Monochromator system have been determined experimentally. The Automated Monochromator was then used for spectral measurements of a platinum lamp. Additionally, the reflectivity curve for a BaS04 coated screen has been measured. Reflectivity measurements indicate a large discrepancy with expected results. Further analysis of the reflectivity experiment is required for conclusive results.

Degree

thesis:*
Name thesis:degree_name
M.S. in Astronautical Engineering
Level thesis:degree_level
Masters
Discipline thesis:degree_discipline
Astronautical Engineering
Department dc:contributor.department
Astronautical Engineering
College dc:contributor.school
Naval Postgraduate School
Grantor dc:publisher
Monterey, California. Naval Postgraduate School
Year dc:date.issued
1992

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Rivers, Thane Damian
Advisor dc:contributor.advisor
  • Cleary, David D.
Committee member dc:contributor.committeemember
  • Davis, Scott

Rights

dc:rights
Statement dc:rights
  • Approved for public release; distribution is unlimited.

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/10945/23866

Chain of custody

source
Harvested from
Naval Postgraduate School
Base URL
calhoun.nps.edu/server/oai/request
Last updated
2026-11-20
Source record
OAI-PMH GetRecord
related terms
citation

Rivers, Thane Damian. Development of an automated scanning monochromator for sensitivity calibration of the MUSTANG instrument.. Masters thesis, Monterey, California. Naval Postgraduate School, 1992. https://hdl.handle.net/10945/23866